Zero filled partial fourier phase contrast MR imaging: in vitro and in vivo assessment.

Gilberto Szarf1, Yoav Dori, Dan Rettmann, Aylin Tekes, Khurram Nasir, Luciano Amado, Thomas K F Foo, David A Bluemke.   

Abstract

PURPOSE: To validate partial Fourier phase contrast magnetic resonance (PC MR) with full number of excitation (NEX) PC MR measurements in vitro and in vivo.
MATERIALS AND METHODS: MR flow measurements were performed using a partial Fourier and a full NEX PC MR sequence in a flow phantom and in 10 popliteal and renal arteries of 10 different healthy volunteers. Average velocity, peak velocity, and flow results were calculated and compared with regression analysis.
RESULTS: Excellent correlations in average velocities (r = 0.99, P < 0.001), peak velocities (r = 0.99, P < 0.001), and flow rates (r = 0.98, P < 0.001) were demonstrated in vitro between the two different acquisitions. For the popliteal arteries there was excellent correlation between peak velocities for both acquisitions (r = 0.98, P < 0.0001); the correlation of average velocity measurements when using all data points in the cardiac cycle for all volunteers was 0.96 (P < 0.001). For the renal arteries the same comparison resulted in a good correlation for average velocity (0.93, P < 0.001) and peak velocity measurements (r = 0.91, P = 0.002), although the correlation coefficient for flow rates was 0.88 (P = 0.004). Blurring of the vessel margins was consistently observed on magnitude images acquired with the partial Fourier method, causing overestimation of the vessel area and some error in the flow measurements.
CONCLUSION: Partial Fourier PC MR is able to provide comparable average and peak velocity values when using 1 NEX PC MRI as a reference.

Mesh:

Year:  2006        PMID: 16315213     DOI: 10.1002/jmri.20472

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  1 in total

1.  Real-time MR thermometry for monitoring HIFU ablations of the liver.

Authors:  Andrew B Holbrook; Juan M Santos; Elena Kaye; Viola Rieke; Kim Butts Pauly
Journal:  Magn Reson Med       Date:  2010-02       Impact factor: 4.668

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.