Targeting accuracy of an image guided gating system for stereotactic body radiotherapy.

Stephen E Tenn1, Timothy D Solberg, Paul M Medin.   

Abstract

Recently, a commercial system capable of x-ray image guided patient positioning and respiratory gated delivery has become available. Here we describe the operational principles of this system and investigate its geometric targeting accuracy under controlled conditions. The system tracks breathing via infrared (IR) detection of reflective markers located on the patient's abdomen. Localization kilovoltage (kV) x-rays are triggered from within the gated delivery window portion of the breathing trace and after positioning, the tumour will cross the linac isocentre during gated delivery. We tested geometric accuracy of this system by localizing and delivering gated fields to a moving phantom. Effects of phantom speed, gating window location, timing errors and phantom rotations on positioning and gating accuracy were investigated. The system delivered gated fields to both a moving and static phantom with equal accuracy. The position of the gating window affects accuracy only to the extent that an asymmetric breathing motion could affect dose distribution within its boundaries. Positioning errors were found to be less then 0.5 +/- 0.2 mm for phantom rotations up to 5 degrees. We found and corrected a synchronization error caused by a faulty x-ray duration setting and detected a 60 +/- 20 ms time delay in our linear accelerator.

Entities:  

Mesh:

Year:  2005        PMID: 16306644     DOI: 10.1088/0031-9155/50/23/002

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  3 in total

1.  An EPID based method for performing high accuracy calibration between an optical external marker tracking device and the LINAC reference frame.

Authors:  Zachary Grelewicz; Hyejoo Kang; Rodney D Wiersma
Journal:  Med Phys       Date:  2012-05       Impact factor: 4.071

2.  Intra-fraction setup variability: IR optical localization vs. X-ray imaging in a hypofractionated patient population.

Authors:  Maria Francesca Spadea; Barbara Tagaste; Marco Riboldi; Eleonora Preve; Daniela Alterio; Gaia Piperno; Cristina Garibaldi; Roberto Orecchia; Antonio Pedotti; Guido Baroni
Journal:  Radiat Oncol       Date:  2011-04-15       Impact factor: 3.481

3.  Development and Validation of a MATLAB Software Program for Decoding the Treatment Errors in Real-time Position Management™ Gating-generated Breathing Trace.

Authors:  Naveen Kumawat; Anil Kumar Shrotriya; Malhotra Singh Heigrujam; Kartikeswar Patro; Satendra Kumar; Anil Kumar Bansal; Ram Kishan Munjal; Anil Kumar Anand
Journal:  J Med Phys       Date:  2020-03-13
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.