Influence of heart rate on the detectability and reproducibility of multislice computed tomography for measuring coronary calcium score using a pulsating calcified mock-vessel in comparison with electron beam tomography.

Nobusada Funabashi, Kikuyo Koide, Naoko Mizuno, Kentaro Nagamura, Shigehiro Ochi, Noriyuki Yanagawa, Yumiko Okamoto, Kimiichi Uno, Issei Komuro.   

Abstract

BACKGROUND: The influence of heart rate on detectability and reproducibility of multislice computed tomography (MSCT) for measuring coronary calcium score was evaluated using pulsating calcified mock-vessels and compared with electron beam tomography (EBT).
MATERIALS AND METHODS: Four calcified mock-vessels with 200-350 HU were made to pulsate at a rate of 40-80 beat/min. Retrospective ECG-gating MSCT (Light Speed Ultra 16) scanning with 0.625 mm slice-thickness was performed twice at each pulsation rate. For comparison, EBT (Imatron C150 XP) was performed with 3 and 1.5 mm beam collimation with prospective ECG-gating and calcium scores were measured.
RESULTS: The comparison revealed that MSCT did not have better reproducibility than EBT, but the calcium scores with MSCT were less influenced by pulsation rates than those with EBT. Especially in mild calcification, the calcium scores decreased with EBT with 3 mm beam collimation with increasing pulsation rate, but the scores were stable in any rate with MSCT.
CONCLUSION: MSCT effectively detects coronary calcification, especially mild calcification, without being influenced by heart rate. This is accomplished by reducing the partial volume effect in the through plane using sub millimeter slice thickness, and using appropriate reconstruction methods, which improve temporal resolution.

Entities:  

Mesh:

Year:  2005        PMID: 16297995     DOI: 10.1016/j.ijcard.2005.08.053

Source DB:  PubMed          Journal:  Int J Cardiol        ISSN: 0167-5273            Impact factor:   4.164


  3 in total

1.  A method for calcium quantification by means of CT coronary angiography using 64-multidetector CT: very high correlation with Agatston and volume scores.

Authors:  Bernhard Glodny; Bettina Helmel; Thomas Trieb; Claudia Schenk; Bernadette Taferner; Verena Unterholzner; Alexander Strasak; Johannes Petersen
Journal:  Eur Radiol       Date:  2009-02-24       Impact factor: 5.315

2.  Small calcified coronary atherosclerotic plaque simulation model: minimal size and attenuation detectable by 64-MDCT and MicroCT.

Authors:  Wisnumurti Kristanto; Peter M A van Ooijen; Jaap M Groen; Rozemarijn Vliegenthart; Matthijs Oudkerk
Journal:  Int J Cardiovasc Imaging       Date:  2011-04-21       Impact factor: 2.357

3.  Dose Reduction in Coronary Artery Calcium Scoring Using Mono-Energetic Images from Reduced Tube Voltage Dual-Source Photon-Counting CT Data: A Dynamic Phantom Study.

Authors:  Niels R van der Werf; Margo van Gent; Ronald Booij; Daniel Bos; Aad van der Lugt; Ricardo P J Budde; Marcel J W Greuter; Marcel van Straten
Journal:  Diagnostics (Basel)       Date:  2021-11-25
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.