Detection of blood coagulation and clot formation using quantitative ultrasonic parameters.

Chih-Chung Huang1, Shyh-Hau Wang, Po-Hsiang Tsui.   

Abstract

The ultrasonic parameters of sound velocity, attenuation and integrated backscatter were applied to detect the process of coagulation and clot formation in porcine blood. Fresh porcine blood containing 15% anticoagulant solution was collected. Blood samples with a hematocrit of 45% were obtained by reconstituting the packed erythrocytes with the separated plasma for ultrasound measurements performed with a 10-MHz focused transducer. A 24-mL aliquot of blood was placed in a container and 12 mL of 0.2 mol/L CaCl2 solution was added to induce clot formation. In each measurement, radio-frequency signals of the blood digitized at 100 MHz were collected for 50 min at a temporal resolution of 1 A-line per s. Results showed that all of the parameters increased within the initial 3 min and, then, immediately decreased dramatically as the CaCl2 solution was added. Subsequently, the sound velocity gradually increased with time and the integrated backscatter and attenuation increased in accordance with blood coagulation until approximately 500 and 2600 s, respectively. The integrated backscatter, attenuation and sound velocity can be divided into different stages, including red cell aggregation, reduction in hematocrit, blood coagulation and clot formation, corresponding to variations in the physical and chemical properties of the blood. The integrated backscatter, attenuation and sound velocity increased because of the changes in blood properties during the process of coagulation and clot formation: by 8.2 dB, 0.65 dB/cm, and 0.6%, respectively. These results provide a feasibility for further applying ultrasonic parameters to in vivo monitor the progress of clotting and thrombosis research.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16286034     DOI: 10.1016/j.ultrasmedbio.2005.07.003

Source DB:  PubMed          Journal:  Ultrasound Med Biol        ISSN: 0301-5629            Impact factor:   2.998


  6 in total

1.  A novel ultrasonic method for evaluation of blood clotting parameters.

Authors:  Vacis Tatarunas; Algirdas Voleisis; Reimondas Sliteris; Rymantas Kazys; Liudas Mazeika; Vaiva Lesauskaite
Journal:  J Med Ultrason (2001)       Date:  2018-03-13       Impact factor: 1.314

2.  Viscosity monitoring during hemodiluted blood coagulation using optical coherence elastography.

Authors:  Xiangqun Xu; Jiang Zhu; Junxiao Yu; Zhongping Chen
Journal:  IEEE J Sel Top Quantum Electron       Date:  2018-05-07       Impact factor: 4.544

3.  SAW-Based Phononic Crystal Microfluidic Sensor-Microscale Realization of Velocimetry Approaches for Integrated Analytical Platform Applications.

Authors:  Aleksandr Oseev; Ralf Lucklum; Mikhail Zubtsov; Marc-Peter Schmidt; Nikolay V Mukhin; Soeren Hirsch
Journal:  Sensors (Basel)       Date:  2017-09-23       Impact factor: 3.576

4.  Quantitative Assessment of First Annular Pulley and Adjacent Tissues Using High-Frequency Ultrasound.

Authors:  Yi-Hsun Lin; Tai-Hua Yang; Shyh-Hau Wang; Fong-Chin Su
Journal:  Sensors (Basel)       Date:  2017-01-07       Impact factor: 3.576

5.  Control of fibrinolytic drug injection via real-time ultrasonic monitoring of blood coagulation.

Authors:  Dmitry A Ivlev; Shakhla N Shirinli; Konstantin G Guria; Svetlana G Uzlova; Georgy Th Guria
Journal:  PLoS One       Date:  2019-02-27       Impact factor: 3.240

Review 6.  Technology Advancements in Blood Coagulation Measurements for Point-of-Care Diagnostic Testing.

Authors:  Mohammad Mohammadi Aria; Ahmet Erten; Ozlem Yalcin
Journal:  Front Bioeng Biotechnol       Date:  2019-12-11
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.