Literature DB >> 16283576

Lower energy endovenous laser ablation of the great saphenous vein with 980 nm diode laser in continuous mode.

Hyun S Kim1, Ikechi J Nwankwo, Kelvin Hong, Patrick S J McElgunn.   

Abstract

PURPOSE: To assess clinical outcomes, complication rates, and unit energy applied using 980 nm diode endovenous laser treatment at 11 watts for symptomatic great saphenous vein (GSV) incompetence and reflux disease.
METHODS: Thirty-four consecutive ablation therapies with a 980 nm diode endovenous laser at 11 watts were studied. The diagnosis of GSV incompetence with reflux was made by clinical evaluation and duplex Doppler examinations. The treated GSVs had a mean diameter of 1.19 cm (range 0.5-2.2 cm). The patients were followed with clinical evaluation and color flow duplex studies up to 18.5 months (mean 12.19 months +/- 4.18).
RESULTS: Using 980 nm diode endovenous laser ablation in continuous mode, 100% technical success was noted. The mean length of GSVs treated was 33.82 cm (range 15-45 cm). The mean energy applied during the treatment was 1,155.81 joules (J) +/- 239.50 (range 545.40-1620 J) for a mean treatment duration of 90.77 sec +/- 21.77. The average laser fiber withdrawal speed was 0.35 cm/sec +/- 0.054. The mean energy applied per length of GSV was 35.16 J/cm +/- 8.43. Energy fluence, calculated separately for each patient, averaged 9.82 J/cm(2) +/- 4.97. At up to 18.5 months follow-up (mean 12.19 months), 0% recanalization was noted; 92% clinical improvement was achieved. There was no major complication. Minor complications included 1 patient with hematoma at the percutaneous venotomy site, 1 patient with thrombophlebitis on superficial tributary varices of the treated GSV, 24% ecchymoses, and 32% self-limiting hypersensitivity/tenderness/"pulling" sensation along the treatment area. One patient developed temporary paresthesia. Four endovenous laser ablation treatments (12%) were followed by adjunctive sclerotherapies for improved cosmetic results.
CONCLUSION: Endovenous laser ablation treatment of GSV using a 980 nm diode laser at 11 watts in continuous mode appears safe and effective. Mean energy applied per treated GSV length of 35.16 J/cm or mean laser fluence of 9.82 J/cm(2) appears adequate, resulting in 0% recanalization and low minor complication rates.

Entities:  

Mesh:

Year:  2006        PMID: 16283576     DOI: 10.1007/s00270-004-0317-9

Source DB:  PubMed          Journal:  Cardiovasc Intervent Radiol        ISSN: 0174-1551            Impact factor:   2.740


  2 in total

1.  Investigation of endovenous laser ablation of varicose veins in vitro using 1.885-μm laser radiation.

Authors:  Alexander N Belyaev; Alexey N Chabushkin; Svetlana A Khrushchalina; Oksana A Kuznetsova; Andrey A Lyapin; Konstantin N Romanov; Polina A Ryabochkina
Journal:  Lasers Med Sci       Date:  2016-02-12       Impact factor: 3.161

Review 2.  [Primary varicosis].

Authors:  T Noppeney; S Rewerk; M Winkler; H Nüllen; H C Schmedt
Journal:  Chirurg       Date:  2007-07       Impact factor: 0.920

  2 in total