Rudolf Marx1, Mutaz Qunaibi, Dieter Christian Wirtz, Fritz Uwe Niethard, Thorsten Mumme.
Abstract
BACKGROUND: One of few persisting problems of cemented total knee arthroplasty (TKA) is aseptic loosening of tibial component due to degradation of the interface between bone cement and metallic tibial shaft component, particularly for surface cemented tibial components. Surface cementation technique has important clinical meaning in case of revision and for avoidance of stress shielding. Degradation of the interface between bone cement and bone may be a secondary effect due to excessive crack formation in bone cement starting at the opposite metallic surface.Entities:
Mesh:
Substances:
Year: 2005 PMID: 16262888 PMCID: PMC1295589 DOI: 10.1186/1475-925X-4-61
Source DB: PubMed Journal: Biomed Eng Online ISSN: 1475-925X Impact factor: 2.819
Figure 3Cracks at the metal (left hand)-cement (right hand) interface made visible using the fluorescence penetration technique.
Figure 2Tibial components ready for surface cementation onto the Technobones (below). Note that the central hole was machined such that there was no "press fit" between the TKA and the Technobones. This ensured maximum stress in the interface.
Figure 1Load geometry for servopneumatically controlled 6kN two-column tabletop tester.
Figure 4Gap at the metal (above)-cement (below) interface.
Cumulative crack lengths for various prostheses
| Silbermatt (m = 2) | 13,999 |
| F22 (m = 8) | 24,693 |
| B60 (m = 2) | 27,351 |
| Coated (m = 8) | 937 |
Relative proportion of gap formation by prosthesis type
| all | 0.8 % | |
| Silbermatt (m = 2) | 1.7 % | |
| F22 (m = 8) | 2.5 % | |
| B60 (m = 2) | 69.0 % |
Figure 5Prosthesis with surface specification B60: Weibull distribution of crack lengths. Modal value π = 120 μm (crack with maximum propability P). Median value of crack length δ = 163 μm (P = 63%; from Fig. 6). Note that the values given in Tab. 3 for B60 are slightly different (170 μm and 133 μm for Median and Modal value, respectively) since those values represent the arithmetic means of characteristic values determined for all prosthesis of this type.
Figure 6Prosthesis with surface specification B60: Weibull plot of crack length. Median value of crack length μ = 163 μm. Modal value of crack length π = 120 μm (from Fig. 5). Modul m = 2,1.
Characteristic Weibull parameters of uncoated prostheses
| PSilbermatt | all | 260 | 175 | 52 |
| PF22 | all | 239 | 205 | 112 |
| PB60 | all | 170 | 133 | 177 |