Static versus prospective gated non-breath hold volumetric MDCT imaging of the lungs.

Osama I Saba1, Deokiee Chon, Kenneth Beck, Geoffrey McLennan, Jered Sieren, Joseph Reinhardt, Eric A Hoffman.   

Abstract

RATIONALE AND
OBJECTIVES: The study's aim is to establish lung-imaging methods that provide for the ability to image the lung under dynamic non-breath hold conditions while providing "virtual breath hold" quantifiable volumetric image data sets. Static breath hold images are used as the gold standard for evaluating these virtual breath hold images in both a phantom and sheep.
MATERIALS AND METHODS: Axial methods for gating image acquisition to multiple points in the respiratory cycle interleaved with incremental table stepping during multidetector-row computed tomographic (MDCT) scanning were developed. Data sets are generated over multiple breaths, providing volume images representative of multiple points within a respiratory cycle. To determine the reproducibility and accuracy of the methods, six anesthetized sheep were studied by means of MDCT in nongated and airway-pressure (P(awy))-gated modes in which P(awy) was 0, 7, and 15 cm H2O.
RESULTS: No significant differences were found between coefficients of variation in air volume measured from repeated static scans (1.74% +/- 1.78%), gated scans: inspiratory (1.2% +/- 0.44%) or expiratory gated (1.39% +/- 0.98%), or between static (1.74% +/- 1.78%) and gated (1.39% +/- 0.98%) scanning at similar P(awy) (P > .1). Measured air volumes were larger from static versus gated scans by 5.85% +/- 3.77% at 7 cm H2O and 4.45% +/- 3.6% at 15 cm H2O of P(awy) (P < .05), consistent with hysteresis. Differences between air volumes at 7 and 15 cm H2O measured from either static or gated scans or that delivered by a super syringe were insignificant (P < .05). Visual accuracy of three-dimensional anatomic geometry was achieved, and landmark certainty was within 1 mm across respiratory cycles.
CONCLUSIONS: A method has been shown that provides for accurate gating to respiratory signals during axial scanning. High-resolution volumetric image data sets are achievable while the scanned subject is breathing. Images are quantitatively similar to breath hold images, with differences likely explained by known pressure-volume hysteresis effects.

Entities:  

Mesh:

Year:  2005        PMID: 16253849      PMCID: PMC1421380          DOI: 10.1016/j.acra.2005.08.014

Source DB:  PubMed          Journal:  Acad Radiol        ISSN: 1076-6332            Impact factor:   3.173


  21 in total

1.  Knowledge-based segmentation of thoracic computed tomography images for assessment of split lung function.

Authors:  M S Brown; J G Goldin; M F McNitt-Gray; L E Greaser; A Sapra; K T Li; J W Sayre; K Martin; D R Aberle
Journal:  Med Phys       Date:  2000-03       Impact factor: 4.071

2.  Subsecond multisection CT of regional pulmonary ventilation.

Authors:  Jehangir K Tajik; Deokiee Chon; Chulho Won; Binh Q Tran; Eric A Hoffman
Journal:  Acad Radiol       Date:  2002-02       Impact factor: 3.173

3.  Respiration-correlated spiral CT: a method of measuring respiratory-induced anatomic motion for radiation treatment planning.

Authors:  E C Ford; G S Mageras; E Yorke; C C Ling
Journal:  Med Phys       Date:  2003-01       Impact factor: 4.071

4.  Terminology for excessive dynamic compression of the airways, tracheobronchial collapse, and related conditions.

Authors:  A H Campbell
Journal:  Am Rev Respir Dis       Date:  1968-03

5.  Quantification of thoracic volumes by three-dimensional imaging.

Authors:  S Krayer; K Rehder; K C Beck; P D Cameron; E P Didier; E A Hoffman
Journal:  J Appl Physiol (1985)       Date:  1987-02

6.  Gas mixing within the acinus of the lung.

Authors:  L A Engel
Journal:  J Appl Physiol Respir Environ Exerc Physiol       Date:  1983-03

7.  CT-based assessment of regional pulmonary microvascular blood flow parameters.

Authors:  Chulho Won; Deokiee Chon; Jehangir Tajik; Binh Q Tran; G Blake Robinswood; Kenneth C Beck; Eric A Hoffman
Journal:  J Appl Physiol (1985)       Date:  2003-02-14

8.  The effect of Pendelluft and dead space on nitrogen clearance: mathematical and experimental models and their application to the study of the distribution of ventilation.

Authors:  I Safonoff; G E Emmanuel
Journal:  J Clin Invest       Date:  1967-10       Impact factor: 14.808

9.  Electrocardiogram-correlated image reconstruction from subsecond spiral computed tomography scans of the heart.

Authors:  M Kachelriess; W A Kalender
Journal:  Med Phys       Date:  1998-12       Impact factor: 4.071

10.  Correlation between the respiratory waveform measured using a respiratory sensor and 3D tumor motion in gated radiotherapy.

Authors:  Yoshikazu Tsunashima; Takeji Sakae; Yoshiyuki Shioyama; Kenji Kagei; Toshiyuki Terunuma; Akihiro Nohtomi; Yasuyuki Akine
Journal:  Int J Radiat Oncol Biol Phys       Date:  2004-11-01       Impact factor: 7.038

View more
  2 in total

Review 1.  Functional imaging: CT and MRI.

Authors:  Edwin J R van Beek; Eric A Hoffman
Journal:  Clin Chest Med       Date:  2008-03       Impact factor: 2.878

2.  Posture-dependent human 3He lung imaging in an open-access MRI system: initial results.

Authors:  Leo L Tsai; Ross W Mair; Chih-Hao Li; Matthew S Rosen; Samuel Patz; Ronald L Walsworth
Journal:  Acad Radiol       Date:  2008-06       Impact factor: 3.173

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.