Literature DB >> 16249497

Relationships between standard automated perimetry, HRT confocal scanning laser ophthalmoscopy, and GDx VCC scanning laser polarimetry.

Nicolaas J Reus1, Hans G Lemij.   

Abstract

PURPOSE: This study was designed to determine and compare the relationships between visual function measured with standard automated perimetry (SAP) and structure, either as neuroretinal rim area measured with confocal scanning laser ophthalmoscopy (CSLO), or as retinal nerve fiber layer thickness determined by scanning laser polarimetry with variable corneal compensation (SLP-VCC).
METHODS: Forty-six healthy subjects and 76 glaucoma patients were examined with SAP, with CSLO by means of the commercially available Heidelberg Retina Tomograph I (HRT), and with SLP-VCC by means of the commercially available GDx VCC. The relationships between SAP, expressed either in the typically used decibel scale or as number of abnormal points in the total deviation probability plot, and CSLO and between SAP and SLP-VCC were described with linear and logarithmic regression analysis for global data and six individual sectors. The relationship between measurements with CSLO and SLP-VCC was fit with linear regression analysis.
RESULTS: The relationships between SAP and CSLO and between SAP and SLP-VCC appeared curvilinear for all sectors except the temporal one between SAP and SLP-VCC. For CSLO, a logarithmic fit was significantly better than a linear one for the global data and in the superotemporal and inferonasal sectors. For SLP-VCC, a curvilinear fit was better for the global data and in the superotemporal, superonasal, and inferonasal sectors. CSLO data correlated linearly with SLP-VCC data in all sectors, except temporally.
CONCLUSIONS: CSLO and SLP-VCC showed a very similar curvilinear relationship with SAP. The observed curvilinear relationships confirm earlier reports that these imaging devices appear to detect glaucomatous loss earlier than SAP.

Entities:  

Mesh:

Year:  2005        PMID: 16249497     DOI: 10.1167/iovs.04-1029

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  16 in total

1.  GDx VCC measurements and blue-on-yellow perimetry in ocular hypertension.

Authors:  Yisheng Zhong; Liping Chen; Yu Cheng; Fangzhi Yu
Journal:  Lasers Med Sci       Date:  2012-02-03       Impact factor: 3.161

2.  Structure/function studies and the effects of memantine in monkeys with experimental glaucoma.

Authors:  B'ann T Gabelt; Carol A Rasmussen; Ozan Y Tektas; Charlene B Y Kim; John C Peterson; T Michael Nork; James N Ver Hoeve; Elke Lütjen-Drecoll; Paul L Kaufman
Journal:  Invest Ophthalmol Vis Sci       Date:  2012-04-30       Impact factor: 4.799

3.  Linearity can account for the similarity among conventional, frequency-doubling, and gabor-based perimetric tests in the glaucomatous macula.

Authors:  Hao Sun; Mitchell W Dul; William H Swanson
Journal:  Optom Vis Sci       Date:  2006-07       Impact factor: 1.973

4.  Combining structural and functional measurements to improve estimates of rates of glaucomatous progression.

Authors:  Felipe A Medeiros; Linda M Zangwill; Christopher A Girkin; Jeffrey M Liebmann; Robert N Weinreb
Journal:  Am J Ophthalmol       Date:  2012-02-07       Impact factor: 5.258

5.  Clinical use of multifocal visual-evoked potentials in a glaucoma practice: a prospective study.

Authors:  Carlos Gustavo De Moraes; Jeffrey M Liebmann; Robert Ritch; Donald C Hood
Journal:  Doc Ophthalmol       Date:  2012-04-05       Impact factor: 2.379

6.  Correlation between blue-on-yellow perimetry and scanning laser polarimetry with variable corneal compensation measurements in primary open-angle glaucoma.

Authors:  Yisheng Zhong; Liping Chen; Yu Cheng; Ping Huang
Journal:  Jpn J Ophthalmol       Date:  2009-12-18       Impact factor: 2.447

7.  Structure-function correlations using scanning laser polarimetry in primary angle-closure glaucoma and primary open-angle glaucoma.

Authors:  Pei-Jung Lee; Catherine Jui-Ling Liu; Robert Wojciechowski; Joan E Bailey-Wilson; Ching-Yu Cheng
Journal:  Am J Ophthalmol       Date:  2010-03-04       Impact factor: 5.258

8.  Incorporating risk factors to improve the assessment of rates of glaucomatous progression.

Authors:  Felipe A Medeiros; Linda M Zangwill; Kaweh Mansouri; Renato Lisboa; Ali Tafreshi; Robert N Weinreb
Journal:  Invest Ophthalmol Vis Sci       Date:  2012-04-24       Impact factor: 4.799

9.  Combining structural and functional measurements to improve detection of glaucoma progression using Bayesian hierarchical models.

Authors:  Felipe A Medeiros; Mauro T Leite; Linda M Zangwill; Robert N Weinreb
Journal:  Invest Ophthalmol Vis Sci       Date:  2011-07-29       Impact factor: 4.799

10.  Combining nerve fiber layer parameters to optimize glaucoma diagnosis with optical coherence tomography.

Authors:  Ake Tzu-Hui Lu; Mingwu Wang; Rohit Varma; Joel S Schuman; David S Greenfield; Scott D Smith; David Huang
Journal:  Ophthalmology       Date:  2008-06-02       Impact factor: 12.079

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.