[Usability of self-tonometry under ambulatory conditions to obtain daily profiles of intraocular pressure at home].

R Vogt1, G I Duncker.   

Abstract

BACKGROUND: Diurnal profiles of intraocular pressure at opening times in the doctor's office do not deliver enough information, especially nightly and early morning measurements are lacking. Self-tonometry using the automatic self-tonometer "Ocuton S" might be suitable to achieve valid 24-hour IOP profiles without the need to stay in a hospital.
MATERIALS AND METHODS: 96 people were taught to use the automatic self-tonometer once. Those who had learnt self-tonometry after this training underwent a diurnal IOP profile for 24 hours in the hospital using Goldmann tonometry and self-tonometry every 4 hours at the same times of day. One to three days later, they created a third IOP profile at home using self-tonometry at the same intervals as at the hospital. The IOP profiles were then tested for equality.
RESULTS: 50 people out of 96 were able to perform self-tonometry after one training session. The IOP measurements of all three IOP profiles did not differ by more than 2 mmHg and thus may be considered comparable. There was no significant difference between daily curves, either.
CONCLUSIONS: Self-tonometry using the automatic self-tonometer "Ocuton S" appears to be suitable to record diurnal profiles of intraocular pressure at home.

Entities:  

Mesh:

Year:  2005        PMID: 16240275     DOI: 10.1055/s-2005-858211

Source DB:  PubMed          Journal:  Klin Monbl Augenheilkd        ISSN: 0023-2165            Impact factor:   0.700


  4 in total

1.  [Self-medication with local anaesthetics by glaucoma patients using teletonometry].

Authors:  G Schäffner; S Antal; C Jürgens; F Tost
Journal:  Ophthalmologe       Date:  2007-12       Impact factor: 1.059

2.  [Do intraocular pressure measurements over 48 h make sense?].

Authors:  R Großjohann; F Tost; B Lorenz
Journal:  Ophthalmologe       Date:  2014-10       Impact factor: 1.059

3.  [Evaluation of the self-tonometer Icare ONE in comparison to Goldmann applanation tonometry].

Authors:  V Witte; Ä Glass; R Beck; R Guthoff
Journal:  Ophthalmologe       Date:  2012-10       Impact factor: 1.059

Review 4.  The Application of a Contact Lens Sensor in Detecting 24-Hour Intraocular Pressure-Related Patterns.

Authors:  Sarah C Xu; Angela C Gauthier; Ji Liu
Journal:  J Ophthalmol       Date:  2016-07-25       Impact factor: 1.909

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.