Determining the apical terminus of root-end resected teeth using three modern apex locators: a comparative ex vivo study.

A ElAyouti1, I Kimionis, A-L Chu, C Löst.   

Abstract

AIM: To assess ex vivo the accuracy of various electronic apex locators in locating the apical terminus of root-end resected teeth.
METHODOLOGY: Ninety extracted human posterior teeth (182 root canals) were prepared to a minimum size of 40 and filled with gutta-percha and sealer. After resection of the apical 3 mm of the root, the root canal filling was removed using HERO rotary instruments. The size of the root canal at the apical terminus after removal of the filling ranged from size 50 to 90. The root canal length to the apical terminus was determined using 3 apex locators (Root ZX, Raypex4 and Apex Pointer). A new mounting model that utilized a micrometer was used to perform the measurements and to visually determine the actual position of the apical terminus. The frequency of locating the apical terminus and the corresponding 95% confidence interval (CI) were calculated. Additionally, the coefficient of repeatability of each apex locator and the limits of inter-operator agreement were determined.
RESULTS: All apex locators showed an acceptable repeatability (0.02-0.03 mm coefficient of repeatability) and narrow limits of inter-operator agreement (+0.07 and -0.07 mm). The accuracy of determining the apical terminus within 1 mm in the root canal was as follows: Root ZX 90% (164/182 root-canals) [95%CI: 86-94%], Raypex4 74% (135/182 root-canals) [95%CI: 68-80%], and Apex Pointer 71% (129/182 root canals) [95%CI: 65-77%]. No over-instrumentation resulted when the Root ZX device was used. In contrast, using the Raypex4 or the Apex Pointer device resulted in over-instrumentation in 8 of 182 root canals (4%).
CONCLUSIONS: Under the conditions of this study all three apex locators were able to detect the apical terminus of root-end resected teeth with an acceptable range. The Root ZX device was the most accurate without over-instrumentation of the root canals.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16218976     DOI: 10.1111/j.1365-2591.2005.01024.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  10 in total

1.  Consistency of electronic measurements of endodontic working length when using multiple devices from the same manufacturer-an in vitro study.

Authors:  Franziska Haupt; M Hülsmann
Journal:  Clin Oral Investig       Date:  2018-03-07       Impact factor: 3.573

2.  The accuracy of two electronic apex locators on effect of preflaring and file size: An in vitro study.

Authors:  V Saritha; H Raghu; Twino H Kumar; Shrishail Totad; Laxmikant Kamatagi; Prahlad A Saraf
Journal:  J Conserv Dent       Date:  2021-07-05

3.  In vitro evaluation of Root ZX and Raypex 6 in teeth with different apical diameters.

Authors:  Ugur Aydin; Emrah Karataslioglu; Fatih Aksoy; Cihan Yildirim
Journal:  J Conserv Dent       Date:  2015 Jan-Feb

4.  A comparative assessment of the accuracies of Raypex 5, Raypex 6, iPex and iPex II electronic apex locators: An in vitro study.

Authors:  Mugem Asli Gurel; Bagdagul Helvacioglu Kivanc; Adil Ekici
Journal:  J Istanb Univ Fac Dent       Date:  2017-01-02

5.  Root ZX Electronic Foramen Locator: An Ex Vivo Study of Its Three Models' Precision and Reproducibility.

Authors:  Bernardo Almeida Aguiar; Rafael Santos Reinaldo; Luciana Maria Arcanjo Frota; Mônica Sampaio do Vale; Bruno Carvalho de Vasconcelos
Journal:  Int J Dent       Date:  2017-03-07

6.  Influence of calcium hydroxide residues after using different irrigants on the accuracy of two electronic apex locators: An in vitro study.

Authors:  Nooshin Sadat Shojaee; Zahra Zaeri; Mohammad Mehdi Shokouhi; Fereshteh Sobhnamayan; Alireza Adl
Journal:  Dent Res J (Isfahan)       Date:  2020-01-21

7.  Influence of Open Apex on Working Length Determination Using Cone-Beam Computed Tomography and Apex Locators: A Comparative In Vitro Study.

Authors:  Swati Srivastava; Hanan M Alharbi; Mai Soliman; Elzahraa Eldwakhly; Manal M Abdelhafeez
Journal:  Biomed Res Int       Date:  2022-07-23       Impact factor: 3.246

8.  The effect of file size on the accuracy of the raypex 5 apex locator: an in vitro study.

Authors:  Shiva Sadeghi; Masoomeh Abolghasemi
Journal:  J Dent Res Dent Clin Dent Prospects       Date:  2008-05-15

9.  Effects of dissolving solutions on the accuracy of an electronic apex locator-integrated endodontic handpiece.

Authors:  Yakup Ustun; Ozgur Uzun; Ozgur Er; Murat Maden; Fatma Yalpı; Burhan Can Canakci
Journal:  ScientificWorldJournal       Date:  2013-11-26

10.  Comparative evaluation of the accuracy of two electronic apex locators in determining the working length in teeth with simulated apical root resorption: An in vitro study.

Authors:  Vidya Saraswathi; Archit Kedia; Tina Puthen Purayil; Vasudev Ballal; Aakriti Saini
Journal:  J Conserv Dent       Date:  2016 Sep-Oct
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.