Impingement of acetabular cups in a hip simulator: comparison of highly cross-linked and conventional polyethylene.

Keith G Holley1, Bridgette D Furman, Omotunde M Babalola, Joseph D Lipman, Douglas E Padgett, Timothy M Wright.   

Abstract

We hypothesized that the reduction in toughness that accompanies elevated levels of cross-linking in ultrahigh-molecular weight polyethylene would lead to increased wear and damage in acetabular cups under conditions of impingement with the neck of the femoral component. Test cups were exposed to 2.8, 10, and 20 Mrad of gamma irradiation and were tested in a simulator, incorporating impingement during every wear cycle. Wear behavior was compared based on wear rate, observations of damage, and the number and morphology of wear particles. The 10-Mrad cups had the lowest rate (15.2 +/- 5.1 mg/Mc), followed by the 2.8-Mrad (40.1 +/- 4.2 mg/Mc) and 20-Mrad (67.9 +/- 24.2 mg/Mc) groups. The wear damage was most severe in the 20-Mrad cups, although all 3 groups showed pitting and delaminating. Wear rates and damage of highly cross-linked polyethylene cups were significantly higher in an impingement wear test than in previous simulator studies.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16214007     DOI: 10.1016/j.arth.2005.04.026

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  8 in total

1.  Hard-on-hard total hip impingement causes extreme contact stress concentrations.

Authors:  Jacob M Elkins; Megan K O'Brien; Nicholas J Stroud; Douglas R Pedersen; John J Callaghan; Thomas D Brown
Journal:  Clin Orthop Relat Res       Date:  2011-02       Impact factor: 4.176

Review 2.  Edge loading has a paradoxical effect on wear in metal-on-polyethylene total hip arthroplasties.

Authors:  William H Harris
Journal:  Clin Orthop Relat Res       Date:  2012-11       Impact factor: 4.176

3.  Are Hooded, Crosslinked Polyethylene Liners Associated with a Reduced Risk of Revision After THA?

Authors:  Adrian Bauze; Somen Agrawal; Alana Cuthbert; Richard de Steiger
Journal:  Clin Orthop Relat Res       Date:  2019-06       Impact factor: 4.176

4.  High stress conditions do not increase wear of thin highly crosslinked UHMWPE.

Authors:  Natalie H Kelly; Amar D Rajadhyaksha; Timothy M Wright; Suzanne A Maher; Geoffrey H Westrich
Journal:  Clin Orthop Relat Res       Date:  2010-02       Impact factor: 4.176

5.  The sagittal stem alignment and the stem version clearly influence the impingement-free range of motion in total hip arthroplasty: a computer model-based analysis.

Authors:  Michael Müller; Georg Duda; Carsten Perka; Stephan Tohtz
Journal:  Int Orthop       Date:  2015-07-02       Impact factor: 3.075

6.  Retrieved highly crosslinked UHMWPE acetabular liners have similar wear damage as conventional UHMWPE.

Authors:  David T Schroder; Natalie H Kelly; Timothy M Wright; Michael L Parks
Journal:  Clin Orthop Relat Res       Date:  2011-02       Impact factor: 4.176

7.  Alumina-on-Polyethylene Bearing Surfaces in Total Hip Arthroplasty.

Authors:  Yup Lee Jung; Shin-Yoon Kim
Journal:  Open Orthop J       Date:  2010-02-11

8.  Severe impingement of lumbar disc replacements increases the functional biological activity of polyethylene wear debris.

Authors:  Ryan M Baxter; Daniel W Macdonald; Steven M Kurtz; Marla J Steinbeck
Journal:  J Bone Joint Surg Am       Date:  2013-06-05       Impact factor: 5.284

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.