Stent fracture after endoluminal repair of a carotid artery pseudoaneurysm.

Jean-Paul P M de Vries1, Rob W Meijer, Jos C van den Berg, Joost M R Meijer, Eric D W M van de Pavoordt.   

Abstract

PURPOSE: To present a case in which a covered stent fractured 7 months after implantation in the internal carotid artery (ICA). CASE REPORT: A 59-year-old man presented with a large traumatic pseudoaneurysm of the left ICA. Seven months after successful treatment with a covered Symbiot stent, clinical complaints recurred because the pseudoaneurysm recanalized due to fracture of the stent at its midsection. Surgical stent explantation was needed, with polytetrafluoroethylene graft reconstruction of the ICA. No complications occurred in the periprocedural period. A control duplex scan 12 months later showed exclusion of the pseudoaneurysm and no graft-related complications.
CONCLUSIONS: This case demonstrates an as yet unreported complication of stent-grafting in a carotid artery.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16212463     DOI: 10.1583/05-1597.1

Source DB:  PubMed          Journal:  J Endovasc Ther        ISSN: 1526-6028            Impact factor:   3.487


  3 in total

1.  The use of flat panel angioCT (DynaCT) for navigation through a deformed and fractured carotid stent.

Authors:  Pasquale Mordasini; Fahmi Al-Senani; Jan Gralla; Dai-Do Do; Caspar Brekenfeld; Gerhard Schroth
Journal:  Neuroradiology       Date:  2009-07-28       Impact factor: 2.804

2.  Delayed cerebral infarction due to stent folding deformation following carotid artery stenting.

Authors:  Kwon-Duk Seo; Kyung-Yul Lee; Byung Moon Kim; Sang Hyun Suh
Journal:  Korean J Radiol       Date:  2014-11-07       Impact factor: 3.500

3.  Fracture of a carotid stent and restenosis of common carotid artery.

Authors:  Wojciech Wróbel; Maciej Lesiak; Anna Sinkiewicz
Journal:  Arch Med Sci       Date:  2013-02-10       Impact factor: 3.318

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.