Tolerance and reliability of wireless pH monitoring in children.

Jay A Hochman1, Jennifer Favaloro-Sabatier.   

Abstract

OBJECTIVES: The purpose of this study was to determine whether the placement of a wireless capsule pH monitoring system improved the reproducibility and patient comfort of pH probe studies in children.
METHODS: The records of 50 children who underwent wireless pH monitoring were retrospectively reviewed. Among this group, 44 children (27 males and 17 females) met inclusion criteria. The average age was 11.8 years, with a range from 6 years to 19 years. Each of these patients had a capsule placed 6 cm above the squamocolumnar junction and underwent pH telemetry for 2 days. In addition, 38 of the 44 families were contacted for follow-up to determine the tolerability of the catheter-free monitoring.
RESULTS: Data analysis revealed that the overall reproducibility of a single 24 hour period was 77%. Studies were considered reproducible if the reflux index was normal (pH <4 for less than 5% of study period) or abnormal on both study days. Using McNemar's exact test, we found no significant difference between the two days (P = 0.11). Ten of 44 patients had conflicting results on day 1 compared with results on day 2. The majority (68%) of patients reported some degree of discomfort during the study; however, this pain was generally mild. Ninety-five percent of parents would be willing to have their child undergo pH monitoring in the future with the wireless pH monitoring.
CONCLUSIONS: Catheter-free prolonged esophageal pH monitoring is feasible in children older than 6 years of age. A lack of consistent reproducibility in sequential 24 hour recordings with this technique concurs with findings using the conventional catheter methodologies. The catheter-free system is often associated with discomfort during the study, but these symptoms were generally well tolerated.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16205507     DOI: 10.1097/01.mpg.0000177312.81071.c8

Source DB:  PubMed          Journal:  J Pediatr Gastroenterol Nutr        ISSN: 0277-2116            Impact factor:   2.839


  5 in total

1.  Reproducibility of 24-hour combined multiple intraluminal impedance (MII) and pH measurements in infants and children. Evaluation of a diagnostic procedure for gastroesophageal reflux disease.

Authors:  Kasper Dalby; Rasmus G Nielsen; Simone Markoew; Søren Kruse-Andersen; Steffen Husby
Journal:  Dig Dis Sci       Date:  2007-04-12       Impact factor: 3.199

2.  Effect of anesthesia on gastroesophageal reflux in children: a study using BRAVO wireless pH study measurements.

Authors:  L Rodriguez; A Morley-Fletcher; A Souza; L Rosengaus; S Nurko
Journal:  Neurogastroenterol Motil       Date:  2015-08-12       Impact factor: 3.598

Review 3.  Gastrointestinal diagnosis using non-white light imaging capsule endoscopy.

Authors:  Gerard Cummins; Benjamin F Cox; Gastone Ciuti; Thineskrishna Anbarasan; Marc P Y Desmulliez; Sandy Cochran; Robert Steele; John N Plevris; Anastasios Koulaouzidis
Journal:  Nat Rev Gastroenterol Hepatol       Date:  2019-07       Impact factor: 46.802

Review 4.  Gastroesophageal reflux disease in 2006. The imperfect diagnosis.

Authors:  John T Boyle
Journal:  Pediatr Radiol       Date:  2006-09

5.  Reliability, Safety and Effectiveness of the Bravo(TM) Capsule: A Catheter-free pH Monitoring System for Evaluation of Gastroesophageal Reflux Disease in Children.

Authors:  Manoochehr Karjoo; Mirza Beg; Shannon Kesselring
Journal:  Middle East J Dig Dis       Date:  2012-01
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.