Evaluation of accuracy in activity calculations for the conjugate view method from Monte Carlo simulated scintillation camera images using experimental data in an anthropomorphic phantom.

Lena Jönsson1, Michael Ljungberg, Sven-Erik Strand.   

Abstract

UNLABELLED: Activity determination from scintillation camera images using the conjugate view method may be inaccurate because of variation in scattered radiation from adjacent organs and activity from overlapping tissues. The aim of this study was to simulate patient scintillation camera images and from these evaluate the accuracy of 2 correction methods. The contribution from overlapping tissue activity was also calculated for some organs.
METHODS: Biokinetic data for (99m)Tc-sestamibi obtained in rats was used as input to simulate scintillation camera images with a voxel-based computer phantom using the Monte Carlo method. The organ activity was calculated using the conjugate view method with either the effective attenuation coefficient method or scatter correction using the triple-energy window (TEW) method combined with attenuation correction with a transmission factor image. Images were simulated with activity in organs one by one to evaluate the accuracy of the 2 correction methods and to evaluate the activity contribution from activity in adjacent or overlapping tissues. To allow comparison with the clinical situation, the total activity distribution from the animal study was used to simulate scintillation camera images at different points in time and the calculated activity was compared with both the input data and some patient data from the literature.
RESULTS: The combination of scatter and attenuation correction gave the most accurate calculated activity, +/-10% of the true activity from the images with activity in one organ at a time. In the images similar to the clinical situation, the kidney activity was overestimated up to a factor of 34, mainly because of excretion of activity through the intestines.
CONCLUSION: The scatter correction using the TEW method in combination with attenuation correction with the measured transmission factor resulted in the most accurate activity determination of the methods used. This study also shows that organ activity data calculated from scintillation camera images may be overestimated by >90% because of activity in overlapping tissues.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16204718

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  11 in total

1.  EANM Dosimetry Committee guidance document: good practice of clinical dosimetry reporting.

Authors:  M Lassmann; C Chiesa; G Flux; M Bardiès
Journal:  Eur J Nucl Med Mol Imaging       Date:  2010-08-27       Impact factor: 9.236

2.  Spatial resolution and image qualities of Zr-89 on Siemens Biograph TruePoint PET/CT.

Authors:  Young Sub Lee; Jin Su Kim; Jung Young Kim; Byung Il Kim; Sang Moo Lim; Hee-Joung Kim
Journal:  Cancer Biother Radiopharm       Date:  2014-12-30       Impact factor: 3.099

3.  Absorbed dose and biologically effective dose in patients with high-risk non-Hodgkin's lymphoma treated with high-activity myeloablative 90Y-ibritumomab tiuxetan (Zevalin).

Authors:  C Chiesa; F Botta; A Coliva; M Maccauro; L Devizzi; A Guidetti; C Carlo-Stella; E Seregni; M A Gianni; E Bombardieri
Journal:  Eur J Nucl Med Mol Imaging       Date:  2009-05-20       Impact factor: 9.236

4.  A gamma camera count rate saturation correction method for whole-body planar imaging.

Authors:  Robert F Hobbs; Sébastien Baechler; Srinivasan Senthamizhchelvan; Andrew R Prideaux; Caroline E Esaias; Melvin Reinhardt; Eric C Frey; David M Loeb; George Sgouros
Journal:  Phys Med Biol       Date:  2010-01-14       Impact factor: 3.609

5.  Changes observed in radionuclide bone scans during and after teriparatide treatment for osteoporosis.

Authors:  Amelia E B Moore; Glen M Blake; Kathleen A Taylor; Valerie A Ruff; Asad E Rana; Xiaohai Wan; Ignac Fogelman
Journal:  Eur J Nucl Med Mol Imaging       Date:  2011-11-05       Impact factor: 9.236

6.  Beyond Palliation: Therapeutic Applications of 153Samarium-EDTMP.

Authors:  Breelyn A Wilky; David M Loeb
Journal:  Clin Exp Pharmacol       Date:  2013-06

7.  Comparison of Accuracy in Calculation of Absorbed Dose to Patients Following Bone Scan with (99m)Tc-Marked Diphosphonates by Two Different Background Correction Methods.

Authors:  Daryoush Shahbazi-Gahrouei; Mehri Damoori; Mohammad Bagher Tavakoli; Masoud Moslehi
Journal:  J Med Signals Sens       Date:  2016 Jan-Mar

8.  Determination of Uterus Absorbed Dose by Patients following Myocardial Perfusion Scan using TLD and Conjugate View Methods.

Authors:  Shanei A; Heydari F; Moslehi M
Journal:  J Biomed Phys Eng       Date:  2019-08-01

9.  Estimation of absorbed dose to the kidneys in patients after treatment with 177Lu-octreotate: comparison between methods based on planar scintigraphy.

Authors:  Maria Larsson; Peter Bernhardt; Johanna B Svensson; Bo Wängberg; Håkan Ahlman; Eva Forssell-Aronsson
Journal:  EJNMMI Res       Date:  2012-09-24       Impact factor: 3.138

10.  Estimation of Organ Activity using Four Different Methods of Background Correction in Conjugate View Method.

Authors:  Ahmad Shanei; Maryam Afshin; Masoud Moslehi; Sedighe Rastaghi
Journal:  J Med Signals Sens       Date:  2015 Oct-Dec
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.