Abstract
A method to determine sildenafil in human plasma involving liquid chromatography-tandem mass spectrometry (LC-MS/MS) has been developed. Sildenafil and the internal standard (I.S.), diazepam, are extracted from human plasma with ether-dichloromethane (3:2, v/v) at basic pH and analyzed by reversed-phase high-performance liquid chromatography (HPLC) using methanol-10mM ammonium acetate pH 7.0 (85:15, v/v) as the mobile phase. Detection by electrospray positive ionization mass spectrometry in the multiple-reaction monitoring mode was linear over the concentration range 0.125-40.0 ng/ml. Intra- and inter-day precision of the assay at four concentrations within this range were 2.5-8.0%. The method was used to evaluate plasma concentration-time profiles in healthy volunteers given an oral dose of 20mg sildenafil as a combination tablet also containing apomorphine.Entities:
Mesh:
Substances:
Year: 2005 PMID: 16203188 DOI: 10.1016/j.jchromb.2005.09.008
Source DB: PubMed Journal: J Chromatogr B Analyt Technol Biomed Life Sci ISSN: 1570-0232 Impact factor: 3.205