Optimum locations of the locking loop and knot in tendon sutures based on the locking Kessler method.

Tadahiko Yotsumoto1, Ryuji Mori, Yuji Uchio.   

Abstract

BACKGROUND: We investigated the factors that influence tensile strength and resistance to gap formation at the repair site of tendon suture (stiffness) by comparing (1) the location of the locking loops and (2) the location of the knot.
METHODS: Transected bovine tendons of the medial gastrocnemius (9-11 x 14-16 mm diameter) were sutured with a modified locking Kessler method with a USP (United States Pharmacopeial convention) 2 polyester multifilament suture (0.500-0.599 mm) and loaded to failure using an Instron 5565 tensiometer. The locking loops were located on either the upper surface facing the operator or on side portions of the tendon. The knot was positioned either close to or far from the tendon stump.
RESULTS: The locations of the locking loops did not influence the tensile strength; however, the stiffness of the side loop suture (5.70 +/- 0.09 N/mm) was significantly higher than that of the upper surface loop suture (5.17 +/- 0.10 N/mm). Regarding the knot location, the tensile strength with the knot far from the tendon stump (195.1 +/- 4.8 N) was significantly higher than that with the knot close to the tendon stump (169.0 +/- 3.6 N), although the stiffness was unchanged by differences in knot location.
CONCLUSIONS: These data suggested that a greater tensile strength with less of a gap is obtained by (1) forming locking loops in the side portion of the tendon, and (2) forming knots far from the tendon stump.

Entities:  

Mesh:

Year:  2005        PMID: 16193365     DOI: 10.1007/s00776-005-0929-1

Source DB:  PubMed          Journal:  J Orthop Sci        ISSN: 0949-2658            Impact factor:   1.601


  3 in total

1.  Biomechanical evaluation of a novel suturing scheme for grafting load-bearing collagen scaffolds for rotator cuff repair.

Authors:  Anowarul Islam; Michael S Bohl; Andrew G Tsai; Mousa Younesi; Robert Gillespie; Ozan Akkus
Journal:  Clin Biomech (Bristol, Avon)       Date:  2015-05-16       Impact factor: 2.063

2.  Comparison of Transverse Intraosseous Loop Technique and Pull Out Suture for Reinsertion of the Flexor Digitorum Profundus tendon. A Retrospective Study.

Authors:  István Zoltán Rigó; Magne Røkkum
Journal:  J Hand Microsurg       Date:  2013-07-31

3.  Evaluating different closed loop graft preparation technique for tibial suspensory fixation in ACL reconstruction using TightRope™.

Authors:  Takahisa Sasho; Toshihide Sasaki; Hiroko Hoshi; Ryuichiro Akagi; Takahiro Enomoto; Yusuke Sato; Ryosuke Nakagawa; Masamichi Tahara; Satoshi Yamaguchi
Journal:  Asia Pac J Sports Med Arthrosc Rehabil Technol       Date:  2018-03-07
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.