Ryoichi Saito1, Yoshiki Misawa1, Kyoji Moriya1, Kazuhiko Koike1, Kimiko Ubukata1, Noboru Okamura1.
Abstract
A loop-mediated isothermal amplification (LAMP) assay for the rapid detection of Mycoplasma pneumoniae was developed and evaluated. The assay specifically amplified only M. pneumoniae sequences, and no cross-reactivity was observed for other Mycoplasma species or respiratory bacterial species. The detection limit for this assay was found to be 2 x 10(2) copies, corresponding to 2-20 colour changing units of M. pneumoniae in 1 h, as observed in a real-time turbidimeter and electrophoretic analysis. The accuracy of the LAMP reaction was confirmed by restriction endonuclease analysis as well as direct sequencing of the amplified product. The assay was applied to 95 nasopharyngeal swab samples collected from patients or from healthy individuals, and compared to a real-time PCR assay in-house. A concordance of 100% was observed between the two assays. The LAMP assay is easy to perform, shows a rapid reaction and is inexpensive. It may therefore be applied in the routine diagnosis of M. pneumoniae infection in the clinical laboratory.Entities:
Mesh:
Substances:
Year: 2005 PMID: 16192434 DOI: 10.1099/jmm.0.46071-0
Source DB: PubMed Journal: J Med Microbiol ISSN: 0022-2615 Impact factor: 2.472