Chest radiography with a flat-panel detector: image quality with dose reduction after copper filtration.

Okka W Hamer1, Claude B Sirlin, Michael Strotzer, Ingitha Borisch, Niels Zorger, Stefan Feuerbach, Markus Völk.   

Abstract

PURPOSE: To compare image quality and estimated dose for chest radiographs obtained by using a cesium iodide-amorphous silicon flat-panel detector at fixed tube voltage and detector entrance dose with and without additional 0.3-mm copper filtration.
MATERIALS AND METHODS: The study was approved by the institutional ethics committee. All prospectively enrolled patients signed the written consent form. Chest radiographs in two projections were acquired at 125-kVp tube voltage and 2.5-microGy detector entrance dose. The experimental group (38 patients) was imaged with 0.3-mm copper filtration; the control group (38 patients) was imaged without copper filtration. An additional 12 patients were imaged with and without copper filtration and served as paired subject-controls. Three readers blinded to group and clinical data independently evaluated the radiographs for image quality on a digital display system. Twelve variables (six for each radiographic projection) were assigned scores on a seven-point ordinal scale. Scores between experimental and control groups were compared: Logistic regression analysis and Mann-Whitney U test were used for unpaired patients; and Wilcoxon and McNemar test, for paired patients. In all, 72 comparisons were determined (36 [12 variables x three readers] for unpaired patients and 36 for paired patients). In a phantom study, radiation burden of experimental protocol was compared with that of control protocol by using Monte Carlo calculations.
RESULTS: For 70 of 72 comparisons, digital radiographs obtained with copper filtration were of similar image quality as radiographs obtained without copper filtration (P = .123 to P > .99). For two of 72 comparisons, one observer judged the experimental protocol superior to the control protocol (P = .043, P = .046). Patient dose reduction estimated with Monte Carlo calculations was 31%. Use of copper filtration increased exposure times by 48% for posteroanterior views and by 34% for lateral views.
CONCLUSION: Subjectively equivalent chest radiographic image quality was found with estimated 30% dose reduction after addition of 0.3-mm copper filtration with flat-panel cesium iodide-amorphous silicon technology.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16192324     DOI: 10.1148/radiol.2372041738

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  12 in total

1.  Visual grading regression: analysing data from visual grading experiments with regression models.

Authors:  O Smedby; M Fredrikson
Journal:  Br J Radiol       Date:  2010-03-11       Impact factor: 3.039

2.  Optimal beam quality for chest flat panel detector system: realistic phantom study.

Authors:  Chie Kuwahara; Takatoshi Aoki; Nobuhiro Oda; Jun Kawabata; Koichiro Sugimoto; Michiko Kobayashi; Masami Fujii; Yukunori Korogi
Journal:  Eur Radiol       Date:  2019-02-08       Impact factor: 5.315

3.  [Dose reduction and adequate image quality in digital radiography: a contradiction?].

Authors:  S Pötter-Lang; M Dünkelmeyer; M Uffmann
Journal:  Radiologe       Date:  2012-10       Impact factor: 0.635

4.  Dose reduction and image quality improvement of chest radiography by using bone-suppression technique and low tube voltage: a phantom study.

Authors:  Satoshi Takagi; Tatsuya Yaegashi; Masayori Ishikawa
Journal:  Eur Radiol       Date:  2019-08-05       Impact factor: 5.315

5.  Digital radiography of crush thoracic trauma in the Sichuan earthquake.

Authors:  Zhi-Hui Dong; Heng Shao; Tian-Wu Chen; Zhi-Gang Chu; Wen Deng; Si-Shi Tang; Jing Chen; Zhi-Gang Yang
Journal:  World J Radiol       Date:  2011-11-28

6.  Pelvic crush fractures in survivors of the Sichuan earthquake evaluated by digital radiography and multidetector computed tomography.

Authors:  Tian-Wu Chen; Zhi-Gang Yang; Zhi-Hui Dong; Zhi-Gang Chu; Jin Yao; Qi-Ling Wang
Journal:  Skeletal Radiol       Date:  2010-03-17       Impact factor: 2.199

7.  Reduction of Fluoroscopy Time and Radiation Dosage During Catheter Ablation for Atrial Fibrillation.

Authors:  Kenichiro Yamagata; Bashar Aldhoon; Josef Kautzner
Journal:  Arrhythm Electrophysiol Rev       Date:  2016-08

8.  Digital tomosynthesis of the chest: comparison of patient exposure dose and image quality between standard default setting and low dose setting.

Authors:  Hye Sun Hwang; Myung Jin Chung; Kyung Soo Lee
Journal:  Korean J Radiol       Date:  2013-05-02       Impact factor: 3.500

9.  Crush extremity fractures associated with the 2008 Sichuan earthquake: anatomic sites, numbers and statuses evaluated with digital radiography and multidetector computed tomography.

Authors:  Tian-wu Chen; Zhi-gang Yang; Qi-ling Wang; Zhi-hui Dong; Jian-qun Yu; Zhang-pu Zhuang; Chang-ling Hou; Zhen-lin Li
Journal:  Skeletal Radiol       Date:  2009-06-25       Impact factor: 2.199

Review 10.  A paediatric X-ray exposure chart.

Authors:  Stephen P Knight
Journal:  J Med Radiat Sci       Date:  2014-06-09
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.