Literature DB >> 16189974

Development of a force amplitude- and location-sensing device designed to improve the ligament balancing procedure in TKA.

Denis Crottet1, Thomas Maeder, Daniel Fritschy, Hannes Bleuler, Lutz P Nolte, Ion P Pappas.   

Abstract

To improve the ligament balancing procedure during total knee arthroplasty a force-sensing device to intraoperatively measure knee joint forces and moments has been developed. It consists of two sensitive plates, one for each condyle, a tibial base plate and a set of spaces to adapt the device thickness to the patient-specific tibiofemoral gap. Each sensitive plate is equipped with three deformable bridges instrumented with thick-film piezoresistive sensors, which allow accurate measurements of the amplitude and location of the tibiofemoral contact forces. The net varus-valgus moment is then computed to characterize the ligamentous imbalance. The developed device has a measurement range of 0-500 N and an intrinsic accuracy of 0.5% full scale. Experimental trials on a plastic knee joint model and on a cadaver specimen demonstrated the proper function of the device in situ. The results obtained indicated that the novel force-sensing device has an appropriate range of measurement and a strong potential to offer useful quantitative information and effective assistance during the ligament balancing procedure in total knee arthroplasty.

Entities:  

Mesh:

Year:  2005        PMID: 16189974     DOI: 10.1109/TBME.2005.851504

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  6 in total

Review 1.  [The foundations of computer assisted surgery].

Authors:  F Langlotz; L-P Nolte; M Tannast
Journal:  Orthopade       Date:  2006-10       Impact factor: 1.087

2.  Ligament tension in the ACL-deficient knee: assessment of medial and lateral gaps.

Authors:  David Mayman; Christopher Plaskos; Daniel Kendoff; G Wernecke; Andrew D Pearle; Richard Laskin
Journal:  Clin Orthop Relat Res       Date:  2009-02-24       Impact factor: 4.176

3.  Internal-external malalignment of the femoral component in kinematically aligned total knee arthroplasty increases tibial force imbalance but does not change laxities of the tibiofemoral joint.

Authors:  Jeremy Riley; Joshua D Roth; Stephen M Howell; Maury L Hull
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2017-11-27       Impact factor: 4.342

4.  The relevance of ligament balancing in total knee arthroplasty: how important is it? A systematic review of the literature.

Authors:  Sina Babazadeh; James D Stoney; Keith Lim; Peter F M Choong
Journal:  Orthop Rev (Pavia)       Date:  2009-10-10

Review 5.  Sensor-Assisted Total Knee Arthroplasty: A Narrative Review.

Authors:  Cheol Hee Park; Sang Jun Song
Journal:  Clin Orthop Surg       Date:  2021-02-15

6.  Enclosed electronic system for force measurements in knee implants.

Authors:  David Forchelet; Matteo Simoncini; Arash Arami; Arnaud Bertsch; Eric Meurville; Kamiar Aminian; Peter Ryser; Philippe Renaud
Journal:  Sensors (Basel)       Date:  2014-08-14       Impact factor: 3.576

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.