Lung volume assessment for a cross-comparison of two breathing-adapted techniques in radiotherapy.

Luc Simon1, Philippe Giraud, Vincent Servois, Jean-Claude Rosenwald.   

Abstract

PURPOSE: To assess the validity of gated radiotherapy of lung by using a cross-check methodology based on four-dimensional (4D)-computed tomography (CT) exams. Variations of volume of a breathing phantom was used as an indicator. METHODS AND MATERIALS: A balloon was periodically inflated and deflated by a medical ventilator. The volume variation (DeltaV) of the balloon was measured simultaneously by a spirometer, taken as reference, and by contouring 4D-CT series (10 phases) acquired by the real-time position management system (RPM). Similar cross-comparison was performed for 2 lung patients, 1 with free breathing (FB), the other with deep-inspiration breath-hold (DIBH) technique.
RESULTS: During FB, DeltaV measured by the spirometer and from 4D-CT were in good agreement: the mean differences for all phases were 8.1 mL for the balloon and 10.5 mL for a patient-test. End-inspiration lung volume has been shown to be slightly underestimated by the 4D-CT. The discrepancy for DeltaV between DIBH and end-expiration, measured from CT and from spirometer, respectively, was less than 3%.
CONCLUSIONS: Provided that each slice series is correctly associated with the proper breathing phase, 4D-CT allows an accurate assessment of lung volume during the whole breathing cycle (DeltaV error <3% compared with the spirometer signal). Taking the lung volume variation into account is a central issue in the evaluation and control of the toxicity for lung radiation treatments.

Entities:  

Mesh:

Year:  2005        PMID: 16168852     DOI: 10.1016/j.ijrobp.2005.05.020

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  6 in total

1.  Prospective respiratory-triggered 64-slice CT pulmonary angiography for detection of pulmonary embolism--a feasibility study in a porcine model.

Authors:  Frank Oliver G Henes; Michael Groth; Marc Regier; Thorsten Bley; Hans-Dieter Nagel; Gerhard Adam; Philipp G C Begemann
Journal:  Emerg Radiol       Date:  2010-07-04

2.  Breath-hold monitoring and visual feedback for radiotherapy using a charge-coupled device camera and a head-mounted display: system development and feasibility.

Authors:  Tadamasa Yoshitake; Katsumasa Nakamura; Yoshiyuki Shioyama; Satoshi Nomoto; Saiji Ohga; Takashi Toba; Takehiro Shiinoki; Shigeo Anai; Hiromi Terashima; Junji Kishimoto; Hiroshi Honda
Journal:  Radiat Med       Date:  2008-01-31

3.  Combined kV and MV imaging for real-time tracking of implanted fiducial markers.

Authors:  R D Wiersma; Weihua Mao; L Xing
Journal:  Med Phys       Date:  2008-04       Impact factor: 4.071

4.  Lung motion and volume measurement by dynamic 3D MRI using a 128-channel receiver coil.

Authors:  Junichi Tokuda; Melanie Schmitt; Yanping Sun; Samuel Patz; Yi Tang; Carolyn E Mountford; Nobuhiko Hata; Lawrence L Wald; Hiroto Hatabu
Journal:  Acad Radiol       Date:  2009-01       Impact factor: 3.173

5.  Dynamic MLC Tracking Using 4D Lung Tumor Motion Modelling and EPID Feedback.

Authors:  Rostampour N; Jabbari K; Nabavi Sh; Mohammadi M; Esmaeili M
Journal:  J Biomed Phys Eng       Date:  2019-08-01

6.  Initial evaluation of a four-dimensional computed tomography system using a programmable motor.

Authors:  Luc Simon; Philippe Giraud; Vincent Servois; Jean-Claude Rosenwald
Journal:  J Appl Clin Med Phys       Date:  2006-11-28       Impact factor: 2.102

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.