| Literature DB >> 1616598 |
Abstract
A simple method for performing manual dose calculations in asymmetric fields is proposed. The method is based on a generalized central-axis dose calculation equation for which open- and wedged-field off-axis provisions have been made. A calculation form designed to document the calculation and simplify the calculation process is presented. Lastly, the required off-axis dosimetry data obtained from a dual-energy accelerator are shown.Mesh:
Year: 1992 PMID: 1616598 DOI: 10.1016/0958-3947(92)90020-g
Source DB: PubMed Journal: Med Dosim ISSN: 1873-4022 Impact factor: 1.482