Analysis of biomechanical factors affecting stent-graft migration in an abdominal aortic aneurysm model.

Z Li1, C Kleinstreuer.   

Abstract

Focusing on a representative abdominal aortic aneurysm (AAA) with a bifurcating stent-graft (SG), a fluid-structure interaction (FSI) solver with user-supplied programs has been employed to solve for blood flow, AAA/SG deformation, sac pressure and wall stresses, as well as the downward forces acting on the SG. Simulation results indicate that implanting a SG can significantly reduce sac pressure, mechanical stress, pulsatile wall motion, and maximum diameter change in AAAs; hence, it may restore normal blood flow and prevent AAA rupture effectively. The transient SG drag force is similar in trend as the cardiac pressure. Its magnitude depends on multi-factors including blood flow conditions, as well as SG and aneurysm geometries. Specifically, AAA neck angle, iliac bifurcation angle, neck aorta-to-iliac diameter ratio, SG size, and blood waveform play important roles in generating a fluid flow force potentially leading to SG migration. It was found that the drag force can exceed 5N for an AAA with a large neck or iliac angle, wide aortic neck and narrow iliac arteries, large SG size, and/or abnormal blood waveform. Thus, the fixation of self-expandable or balloon-expandable SG contact may be inadequate to withstand the forces of blood flowing through the implant and hence means of extra fixation should be considered. A comprehensive FSI analysis of the coupled SG-AAA dynamics provides physical insight for evaluating the luminal hemodynamics, and maximum AAA-stresses as well as biomechanical factors leading potentially to SG migration.

Entities:  

Mesh:

Year:  2005        PMID: 16153654     DOI: 10.1016/j.jbiomech.2005.07.010

Source DB:  PubMed          Journal:  J Biomech        ISSN: 0021-9290            Impact factor:   2.712


  20 in total

1.  Effect of curvature on displacement forces acting on aortic endografts: a 3-dimensional computational analysis.

Authors:  C Alberto Figueroa; Charles A Taylor; Victoria Yeh; Allen J Chiou; Christopher K Zarins
Journal:  J Endovasc Ther       Date:  2009-06       Impact factor: 3.487

2.  A computational framework for investigating the positional stability of aortic endografts.

Authors:  Anamika Prasad; Nan Xiao; Xiao-Yan Gong; Christopher K Zarins; C Alberto Figueroa
Journal:  Biomech Model Mechanobiol       Date:  2012-11-10

3.  Engineering silicone rubbers for in vitro studies: creating AAA models and ILT analogues with physiological properties.

Authors:  T J Corbett; B J Doyle; A Callanan; M T Walsh; T M McGloughlin
Journal:  J Biomech Eng       Date:  2010-01       Impact factor: 2.097

4.  Original article submission: Platelet stress accumulation analysis to predict thrombogenicity of an artificial kidney.

Authors:  Amanda K W Buck; Steven G Goebel; Mark S Goodin; Nathan J Wright; Joseph J Groszek; Jarrett Moyer; Sukhveer Singh; Danny Bluestein; William H Fissell; Shuvo Roy
Journal:  J Biomech       Date:  2018-01-16       Impact factor: 2.712

5.  Combined In Silico and In Vitro Approach Predicts Low Wall Shear Stress Regions in a Hemofilter that Correlate with Thrombus Formation In Vivo.

Authors:  Amanda K W Buck; Joseph J Groszek; Daniel C Colvin; Sara B Keller; Clark Kensinger; Rachel Forbes; Seth Karp; Phillip Williams; Shuvo Roy; William H Fissell
Journal:  ASAIO J       Date:  2018 Mar/Apr       Impact factor: 2.872

6.  Role of graft oversizing in the fixation strength of barbed endovascular grafts.

Authors:  Jarin A Kratzberg; Jafar Golzarian; Madhavan L Raghavan
Journal:  J Vasc Surg       Date:  2009-06       Impact factor: 4.268

Review 7.  A literature review of the numerical analysis of abdominal aortic aneurysms treated with endovascular stent grafts.

Authors:  David Roy; Claude Kauffmann; Sébastien Delorme; Sophie Lerouge; Guy Cloutier; Gilles Soulez
Journal:  Comput Math Methods Med       Date:  2012-09-06       Impact factor: 2.238

8.  Functional Evaluation of Embedded Modular Single-Branched Stent Graft: Application to Type B Aortic Dissection With Aberrant Right Subclavian Artery.

Authors:  Xuehuan Zhang; Duanduan Chen; Mingwei Wu; Huiwu Dong; Zhengdong Wan; Heyue Jia; Shichao Liang; Jun Shao; Jun Zheng; Shangdong Xu; Jiang Xiong; Wei Guo
Journal:  Front Cardiovasc Med       Date:  2022-05-02

9.  Fenestrated stent graft repair of abdominal aortic aneurysm: hemodynamic analysis of the effect of fenestrated stents on the renal arteries.

Authors:  Zhonghua Sun; Thanapong Chaichana
Journal:  Korean J Radiol       Date:  2009-12-28       Impact factor: 3.500

10.  Fluid-structure interaction of a patient-specific abdominal aortic aneurysm treated with an endovascular stent-graft.

Authors:  David S Molony; Anthony Callanan; Eamon G Kavanagh; Michael T Walsh; Tim M McGloughlin
Journal:  Biomed Eng Online       Date:  2009-10-06       Impact factor: 2.819

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.