| Literature DB >> 16129745 |
B M Wroblewski1, P D Siney, P A Fleming.
Abstract
We report the results of our continued review of 11 total hip arthroplasties using 22.225 mm alumina ceramic femoral heads on a Charnley flanged stem, articulating with chemically cross-linked polyethylene. There was an initial bedding-in of up to 0.41 mm at the articular surface in the first two years. This had not progressed further, at a minimum follow-up of 15 years. Radiographically no femoral or acetabular component showed loosening or osteolysis.Entities:
Mesh:
Substances:
Year: 2005 PMID: 16129745 DOI: 10.1302/0301-620X.87B9.16265
Source DB: PubMed Journal: J Bone Joint Surg Br ISSN: 0301-620X