The effects of skin temperature and testing site on blood glucose measurements taken by a modern blood glucose monitoring device.

A Haupt1, B Berg, P Paschen, M Dreyer, H-U Häring, J Smedegaard, S Matthaei.   

Abstract

BACKGROUND: Modern blood glucose (BG) monitoring devices (e.g., InDuo [LifeScan, Inc., Milpitas, CA]) require very low blood volumes, allowing for testing at sites other than the traditional fingertip, but the reliability of such testing has not been fully elucidated. The aim of this randomized study was to compare the effects of cold/warm skin temperature combined with alternative site (forearm) testing versus conventional fingertip measurements on fasting and postprandial conditions.
MATERIALS AND METHODS: The study population consisted of 19 patients who had previously used InDuo for 6 weeks. Four simultaneous (within 1 min) BG readings (left and right forearm and fingertips) were obtained from each patient 15, 10, and 5 min before eating. Ten minutes before eating, the patient immersed one arm in cold water (15.5 degrees C) and the other in warm water (35 degrees C). At time 0 min arms were removed from water baths, and the patient was offered a standard meal (duration 15 min). Arms were again immersed in water baths, and BG was measured from the same locations 20 min after eating and at subsequent 15-min intervals for 185 min. The effects of site testing and temperature were assessed in this period by identifying maximum BG concentrations (C (max)) and time to C (max) (T (max)).
RESULTS: Significantly lower Cmax values were observed for (1) cold forearm versus cold fingertip (mean Delta 28.6 mg/dL, P < 0.001), (2) warm forearm versus warm fingertip (mean Delta 12 mg/dL, P = 0.028), (3) cold fingertip versus warm fingertip (mean Delta 17.2 mg/dL, P = 0.002), and (4) cold forearm versus warm forearm (mean Delta 33.7 mg/dL, P < 0.001). Significantly longer Tmax values were reported for cold forearm versus warm forearm (mean Delta 22.4 min, P < 0.001) and cold forearm versus cold fingertip (mean Delta 20 min, P < 0.001).
CONCLUSIONS: These results demonstrate that cold skin and forearm conditions significantly underestimate BG concentrations and delay T(max) compared with warm skin and fingertip measurements.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16120031     DOI: 10.1089/dia.2005.7.597

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  6 in total

Review 1.  Geoenvironmental diabetology.

Authors:  Curtiss B Cook; Kay E Wellik; Margaret Fowke
Journal:  J Diabetes Sci Technol       Date:  2011-07-01

Review 2.  Factors affecting blood glucose monitoring: sources of errors in measurement.

Authors:  Barry H Ginsberg
Journal:  J Diabetes Sci Technol       Date:  2009-07-01

3.  The need for continuous blood glucose monitoring in the intensive care unit.

Authors:  Ram Weiss; Isaac Lazar
Journal:  J Diabetes Sci Technol       Date:  2007-05

4.  Glucose monitoring after fruit peeling: pseudohyperglycemia when neglecting hand washing before fingertip blood sampling: wash your hands with tap water before you check blood glucose level.

Authors:  Takahisa Hirose; Tomoya Mita; Yoshio Fujitani; Ryuzo Kawamori; Hirotaka Watada
Journal:  Diabetes Care       Date:  2011-01-31       Impact factor: 19.112

5.  Comparison of Accu Chek Inform II point-of-care test blood glucose meter with Hexokinase Plasma method for a diabetes mellitus population during surgery under general anesthesia.

Authors:  B Lascaris; H G Freling; M A Edens; M J Fokkert; C G Olthof; R J Slingerland
Journal:  J Clin Monit Comput       Date:  2021-01-26       Impact factor: 1.977

6.  Self-monitoring of blood glucose during pregnancy: indications and limitations.

Authors:  Carlos Antonio Negrato; Lenita Zajdenverg
Journal:  Diabetol Metab Syndr       Date:  2012-12-22       Impact factor: 3.320

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.