Literature DB >> 16117777

In vitro accuracy of a novel registration and targeting technique for image-guided template production.

Gerlig Widmann1, Roland Widmann, Ekkehard Widmann, Werner Jaschke, Reto Josef Bale.   

Abstract

OBJECTIVES: The objective of this study was to evaluate the accuracy of a novel registration and targeting technique for image-guided template production (IGTP) in a preliminary phantom study.
MATERIAL AND METHODS: Registration of four standard dental stone casts with integrated target pellets to the corresponding computed tomography (CT) data was performed via a vacuum mouthpiece and an external reference frame (Medical Intelligence GmbH, Germany). Using the Treon navigation system (Medtronic Inc., Minneapolis, MN, USA) a surgical path with the entry in the centre of the dental crown and the target in the centre of the target pellet was planned on the CT data. An aiming device was adjusted according to the planned trajectory and guided drillings into the dental stone casts. The accuracy was evaluated on postoperative 3D-CT data.
RESULTS: The mean fiducial registration error as given by the registration software was 0.4 mm. One hundred and twelve navigated drillings showed a mean accuracy [xy] of 0.42+/-0.26 mm (maximum 1 mm). For the z-axis, a mean accuracy [z] of 0.25+/-0.12 mm (maximum 0.6 mm) was found.
CONCLUSIONS: Comparing the presented registration technique to existing registration methods in IGTP and burr tracking, no radiographic and registration templates are needed. The procedure is easy and requires only minimal effort. Navigation-controlled drillings could be performed with an accuracy that approaches the intrinsic navigation system's accuracy, a fact that warrants its use for surgical template production. Further accuracy studies of template-guided drillings are necessary before the presented registration technique can be implemented for patient treatment.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16117777     DOI: 10.1111/j.1600-0501.2005.01131.x

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  6 in total

1.  Comparative study of application accuracy of two frameless neuronavigation systems: experimental error assessment quantifying registration methods and clinically influencing factors.

Authors:  Dimitrios Paraskevopoulos; Andreas Unterberg; Roland Metzner; Jens Dreyhaupt; Georg Eggers; Christian Rainer Wirtz
Journal:  Neurosurg Rev       Date:  2011-01-19       Impact factor: 3.042

2.  Automated dental implantation using image-guided robotics: registration results.

Authors:  Xiaoyan Sun; Frederic D McKenzie; Sebastian Bawab; Jiang Li; Yongki Yoon; Jen-K Huang
Journal:  Int J Comput Assist Radiol Surg       Date:  2011-01-08       Impact factor: 2.924

3.  [Modified procedure for patient registration for navigation control instruments in ENT surgery].

Authors:  G Strauss; C Spitzer; E Dittrich; M Hofer; M Strauss; T Lüth
Journal:  HNO       Date:  2009-02       Impact factor: 1.284

4.  Image-guided surgery and medical robotics in the cranial area.

Authors:  G Widmann
Journal:  Biomed Imaging Interv J       Date:  2007-01-01

5.  Evaluation of Resins for Stereolithographic 3D-Printed Surgical Guides: The Response of L929 Cells and Human Gingival Fibroblasts.

Authors:  Christoph Kurzmann; Klara Janjić; Hassan Shokoohi-Tabrizi; Michael Edelmayer; Manuela Pensch; Andreas Moritz; Hermann Agis
Journal:  Biomed Res Int       Date:  2017-10-22       Impact factor: 3.411

6.  In-vitro study on the accuracy of a simple-design CT-guided stent for dental implants.

Authors:  Young-June Huh; Bo-Ram Choi; Kyung-Hoe Huh; Won-Jin Yi; Min-Suk Heo; Sam-Sun Lee; Soon-Chul Choi
Journal:  Imaging Sci Dent       Date:  2012-09-21
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.