The efficacy of a new portable sequential compression device (SCD Express) in preventing venous stasis.

Stavros K Kakkos1, Maura Griffin, George Geroulakos, Andrew N Nicolaides.   

Abstract

OBJECTIVE: It has been previously shown that the SCD Response Compression System, by sensing the postcompression refill time of the lower limbs, delivers more compression cycles over time, resulting in as much as a 76% increase in the total volume of blood expelled per hour. Extended indications for pneumatic compression have necessitated the introduction of portable devices. The aim of our study was to test the hemodynamic effectiveness of a new portable sequential compression system (the SCD Express), which has the ability to detect the individual refill time of the two lower limbs separately.
METHODS: This was an open, controlled trial with 30 normal volunteers. The new SCD Express was compared with the SCD Response Compression System in the supine and semirecumbent positions. The refilling time sensed by the device was compared with that determined from velocity recordings of the superficial femoral vein using duplex ultrasonography. Baseline and augmented flow velocity and volume flow, including the total volume of blood expelled per hour during compression with the SCD Express, were compared with those produced by the SCD Response compression system in the same volunteers and positions.
RESULTS: Both devices significantly increased venous flow velocity as much as 2.26 times baseline in supine position and 2.67 times baseline in semirecumbent position (all P < .001). There was a linear relationship between duplex ultrasonography-derived refill time and the SCD Express-derived refill time in both the supine (r = 0.39, P = .03) and semirecumbent (r = 0.71, P < .001) positions but not with the SCD Response. Refill time measured by the SCD Express device was significantly shorter and the cycle rate higher in comparison with the SCD Response in both positions. The single-cycle flow velocity and volume flow parameters generated by the two devices were similar in both positions. However, median (interquartile range) total volume of blood expelled per hour was slightly higher with the SCD Express device in the supine position (7206 mL/h [range, 5042-8437] vs 6712 mL/h [4941-10,676]; P = .85) and semirecumbent position (4588 mL/h [range, 3721-6252] vs 4262 mL/h [3520-5831]; P = .22). Peak volume of blood expelled per hour by the SCD Express device in the semirecumbent position was significantly increased by 10% in comparison with the SCD Response (P = .03).
CONCLUSIONS: Flow velocity and volume flow enhancement by the SCD Response and SCD Express were essentially similar. The latter, a portable device with optional battery power that detects the individual refill time of the lower limbs separately, is anticipated to be associated with improved overall compliance and therefore optimized thromboprophylaxis. Studies testing its potential for improved efficacy in preventing deep vein thrombosis are justified.

Entities:  

Mesh:

Year:  2005        PMID: 16102630     DOI: 10.1016/j.jvs.2005.03.031

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  9 in total

1.  What influence does intermittent pneumatic compression of the lower limbs intraoperatively have on core hypothermia?

Authors:  Jin Huh; Yong-Beom Cho; Mi-Kyung Yang; Yeon-Kyeong Yoo; Duk-Kyung Kim
Journal:  Surg Endosc       Date:  2013-01-09       Impact factor: 4.584

2.  Intermittent compression of the calf muscle as a countermeasure to protect blood pressure and brain blood flow in upright posture in older adults.

Authors:  K A Zuj; E T Hedge; J D Milligan; S D Peterson; R L Hughson
Journal:  Eur J Appl Physiol       Date:  2021-01-02       Impact factor: 3.078

Review 3.  Combined intermittent pneumatic leg compression and pharmacological prophylaxis for prevention of venous thromboembolism.

Authors:  Stavros Kakkos; George Kirkilesis; Joseph A Caprini; George Geroulakos; Andrew Nicolaides; Gerard Stansby; Daniel J Reddy
Journal:  Cochrane Database Syst Rev       Date:  2022-01-28

4.  Comparison of clinical and physiological efficacies of different intermittent sequential pneumatic compression devices in preventing deep vein thrombosis: a prospective randomized study.

Authors:  Ki Hyoung Koo; Jae-Sung Choi; Ji Hyun Ahn; Jae Hyun Kwon; Keun-Tae Cho
Journal:  Clin Orthop Surg       Date:  2014-11-10

Review 5.  Reduced cardiac function and risk of venous thromboembolism in Asian countries.

Authors:  Ruiqi Zhu; Liang Tang; Yu Hu
Journal:  Thromb J       Date:  2017-04-24

6.  Incidence of deep venous thrombosis following periacetabular and derotational femoral osteotomy: a case for mechanical prophylaxis.

Authors:  Matthew J Kraeutler; Sivashanmugam Raju; Tigran Garabekyan; Omer Mei-Dan
Journal:  J Hip Preserv Surg       Date:  2018-03-05

Review 7.  Combined intermittent pneumatic leg compression and pharmacological prophylaxis for prevention of venous thromboembolism.

Authors:  Stavros K Kakkos; Joseph A Caprini; George Geroulakos; Andrew N Nicolaides; Gerard Stansby; Daniel J Reddy; Ioannis Ntouvas
Journal:  Cochrane Database Syst Rev       Date:  2016-09-07

8.  The use of a novel in-bed active Leg Exercise Apparatus (LEX) for increasing venous blood flow.

Authors:  Kenta Tanaka; Hiroshi Kamada; Yukiyo Shimizu; Shizu Aikawa; Tomofumi Nishino; Naoyuki Ochiai; Masataka Sakane; Masashi Yamazaki
Journal:  J Rural Med       Date:  2016-06-16

9.  Interpretation of the PREVENT study findings on the adjunctive role of intermittent pneumatic compression to prevent venous thromboembolism.

Authors:  Stavros K Kakkos; Andrew N Nicolaides; Joseph A Caprini
Journal:  Ann Transl Med       Date:  2020-06
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.