| Literature DB >> 1609585 |
A Lindstrand1, A Stenström, S Lewold.
Abstract
The prospective Swedish Knee Arthroplasty Project contained 772 PCA, 1564 Marmor and 1441 St Georg unicompartment prostheses, implanted for gonarthrosis during the period December 1983-July 1990. Follow-up was performed in September 1991. There was a significant difference in the cumulative revision rate after only 2 years, increasing to 15 percent for the PCA, compared with 5-7 percent for the two other types. The difference between PCA and Marmor/St Georg was loosening of the femoral component. Major polyethylene wear was noted in a quarter of the revised PCA tibial components. The wear pattern of Marmor/St Georg was not analyzed in this study.Entities:
Mesh:
Substances:
Year: 1992 PMID: 1609585 DOI: 10.3109/17453679209154777
Source DB: PubMed Journal: Acta Orthop Scand ISSN: 0001-6470