Comparison of dynamic digital pupillometry and static measurements of pupil size in determining scotopic pupil size before refractive surgery.

Sanj S Wickremasinghe1, Guy T Smith, Julian D Stevens.   

Abstract

PURPOSE: To compare 2 digital infrared pupillometers.
SETTING: Anterior Segment Service, Moorfields Eye Hospital, London, England.
METHODS: Eighty-eight eyes of 44 healthy subjects were recruited into the study. The scotopic pupil size was measured with 2 instruments, a digital pupillometer (Procyon) and a Hartman-Shack wavefront aberrometer (Visx WaveScan). Agreement between the instruments was assessed.
RESULTS: There were 21 men and 23 women with a mean age 38.1 years +/- 9.2 (SD) (range 23 to 62 years). The mean scotopic pupil diameter was significantly larger with the WaveScan (6.61 +/- 0.92 mm) than with the Procyon pupillometer (6.40 +/- 0.90 mm), P<.001.
CONCLUSIONS: There was good agreement in measurement of scotopic pupil size between the two instruments. The larger pupil diameter found with the Visx WaveScan than with the Procyon digital pupillometer probably represents the different illumination level used by each instrument, the effects of accommodation, and target fogging.

Entities:  

Mesh:

Year:  2005        PMID: 16039493     DOI: 10.1016/j.jcrs.2004.10.049

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  6 in total

1.  Corneal white-to-white distance and mesopic pupil diameter.

Authors:  Hasan Basri Cakmak; Nurullah Cagil; Huseyin Simavli; Sabri Raza
Journal:  Int J Ophthalmol       Date:  2012-08-18       Impact factor: 1.779

2.  Digital Pupillometry in Normal Subjects.

Authors:  Annekatrin Rickmann; Maria Waizel; Sara Kazerounian; Peter Szurman; Helmut Wilhelm; Karl T Boden
Journal:  Neuroophthalmology       Date:  2016-10-03

3.  Dysfunction of the pupillary light reflex in experimental autoimmune autonomic ganglionopathy.

Authors:  Shalini Mukherjee; Steven Vernino
Journal:  Auton Neurosci       Date:  2007-07-05       Impact factor: 3.145

4.  Efficacy of digital pupillometry for diagnosis of Horner syndrome.

Authors:  Yung Ju Yoo; Hee Kyung Yang; Jeong-Min Hwang
Journal:  PLoS One       Date:  2017-06-02       Impact factor: 3.240

5.  Differences in pupillary light reflex between optic neuritis and ischemic optic neuropathy.

Authors:  Yung Ju Yoo; Jeong-Min Hwang; Hee Kyung Yang
Journal:  PLoS One       Date:  2017-10-19       Impact factor: 3.240

6.  Intrasession repeatability of pupil size measurements under different light levels provided by a multidiagnostic device in healthy eyes.

Authors:  David P Piñero; Dolores de Fez; Inmaculada Cabezos; Alberto López-Navarro; María T Caballero; Vicente J Camps
Journal:  BMC Ophthalmol       Date:  2020-08-31       Impact factor: 2.209

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.