Oscillometric blood pressure devices and simulators: measurements of repeatability and differences between models.

A J Sims1, C A Reay, D R Bousfield, J A Menes, A Murray.   

Abstract

PRIMARY
OBJECTIVE: To measure the repeatability and pressure pulse envelope of simulators used for testing oscillometric non-invasive blood pressure (NIBP) devices; to study the effect of different envelopes on NIBP devices, and to measure the difference between NIBP devices due to different oscillometric algorithms.
METHODS: Three different models of NIBP simulator and 18 different patient monitors with NIBP function were studied. We developed a pressure measurement system (accuracy 0.048?mmHg) to measure the repeatability of simulators. The effect of changing the envelope was measured by using three simulators with one NIBP device. Differences between 18 NIBP devices were measured using one simulator at seven blood pressure settings. MAIN OUTCOMES AND
RESULTS: Simulators generate repeatable pressure pulse envelopes (< 0.2 mmHg) but the magnitude and shape depends on the model of simulator. Oscillometric NIBP devices are highly repeatable (< 2 mmHg) when presented with a repeatable pressure pulse envelope, but different devices employ different algorithms and give different results. For a simulated standard blood pressure setting of 120/80 mmHg, estimates of systolic pressure ranged from 112.6 to 126.6 mmHg (sd of 3.0 mmHg), and diastolic pressure ranged from 74.8 to 86.9 mmHg (sd of 3.5 mmHg).
CONCLUSIONS: Simulators and NIBP devices are sufficiently repeatable for clinical use, but further systematic clinical studies are required to better characterize the pressure pulse envelope for different patient groups.

Entities:  

Mesh:

Year:  2005        PMID: 16019880     DOI: 10.1080/03091900500067983

Source DB:  PubMed          Journal:  J Med Eng Technol        ISSN: 0309-1902


  2 in total

Review 1.  A procedure for evaluation of non-invasive blood pressure simulators.

Authors:  Gregor Gersak; Ales Zemva; Janko Drnovsek
Journal:  Med Biol Eng Comput       Date:  2009-10-13       Impact factor: 2.602

2.  A convenient method to verify the accuracy of oscillometric blood pressure monitors by the auscultatory method: A smartphone-based app.

Authors:  Zhi Zhang; Weichun Xi; Bingjiang Wang; Guang Chu; Fang Wang
Journal:  J Clin Hypertens (Greenwich)       Date:  2018-12-20       Impact factor: 3.738