| Literature DB >> 16018819 |
Rasik B Vajpayee1, Kamna Verma, Rajesh Sinha, Jeewan S Titiyal, R M Pandey, Namrata Sharma.
Abstract
BACKGROUND: Various ophthalmic viscosurgical devices (OVD) are used to perform phacoemulsification and other intraocular surgeries. We performed a study to compare the efficacy and safety of three ophthalmic viscosurgical devices that are routinely used in phacoemulsification.Entities:
Mesh:
Substances:
Year: 2005 PMID: 16018819 PMCID: PMC1180454 DOI: 10.1186/1471-2415-5-17
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.209
Pre operative and Post operative characteristics of the patients
| Age | 69.6 ± 9.2 | 65.8 ± 7.8 | 70.8 ± 9.9 | 1.58 | NS |
| ACD | 2.6 ± 0.20 | 2.8 ± 0.37 | 2.59 ± 0.45 | 1.97 | NS |
| UCVA | |||||
| Pre-operative | 0.12 ± 0.14 | 0.09 ± 0.13 | 0.11 ± 0.9 | 0.18 | NS |
| Day one | 0.28 ± 19a | 0.39 ± 23c | 0.37 + 0.27b | 1.22 | NS |
| Three month | 0.51 ± 0.16c | 0.61 ± 0.20c | 0.61 ± 0.20c | 1.51 | NS |
| Repeated measure within group ANOVA | F = 15.0 | F = 28.29 | F = 23.5 | ||
| BCVA | |||||
| Pre-operative | 0.18 ± 0.11 | 0.17 ± 0.19 | 0.18 ± 0.13 | 0.04 | NS |
| Day one | 0.46 ± 0.23b | 0.69 ± 0.26c | 0.55 ± 0.31a | 1.70 | NS |
| Three month | 0.69 ± 0.24c | 0.88 ± 0.18c | 0.79 ± 0.23c | 2.73 | NS |
| Repeated measure within group ANOVA | F = 34.17 | F = 43.8 | F = 23.5 | ||
| Corneal thickness | |||||
| Pre-operative | 560.7 ± 91.5 | 567.5 ± 44.07 | 576.0 ± 50.2 | 0.25 | NS |
| Day one | 645.8 ± 94.1c | 665.5 ± 141.9b | 670.6 ± 117.1b | 0.32 | NS |
| Three month | 573.4 ± 73.6 | 586.1 ± 29.0 | 586.7 ± 36.0 | 1.50 | NS |
| Repeated measure within group ANOVA | F = 17.5 | F = 11.4 | F = 11.19 | ||
| Endothelial cell count. | |||||
| Pre-operative | 2311.73 ± 288.0 | 2359.78 ± 383.2 | 2526.6 ± 305.4 | 2.36 | NS |
| Day one | 2021.20 ± 201.0 | 2035.05 ± 377.0 | 2196.4 ± 378.2 | 0.11 | NS |
| Three month | 1973.00 ± 537.0 | 1959.00 ± 503.0 | 2132.0 ± 482.0 | 0.82 | NS |
| Repeated measure within group ANOVA | F = 4.2 | F = 2.20 | F = 17.5 |
a: p < 0.05, b: <0.01, c: <0.001 as compared to preoperative parameter
UCVA = Uncorrected visual acuity
BCVA = Best corrected visual acuity
ANOVA = Analysis of Variance
NS = Not Significant
Intra-operative oarameters
| Mean phaco energy (%) | 0.28 ± 0.06 | 0.31 ± 0.06 | 0.38 ± 0.09 | 6.75 | 0.0024 | I Vs III: p = 0.03 |
| Mean phaco time (min) | 2.0 ± 1.0 | 1.8 ± 0.8 | 2.1 ± 0.7 | 0.026 | -- | -- |
| Effective phaco time (sec) | 33.8 ± 17.7 | 36.1 ± 20.4 | 49.5 ± 23.1 | 3.31 | -- | -- |
| Mean total surgical time (min) | 12.9 ± 3.3 | 15.2 ± 3.5 | 16.5 ± 3.6 | 5.06 | 1 Vs III: p = 0.011 | |
| OVD removal time (sec) | 66.6 ± 11.2 | 45.1 ± 9.0 | 55.47 ± 6.6 | 26 | 0.0001 | I Vs II: p = 0.0001 |
OVD = Ophthalmic viscosurgical device
min = minutes
sec = seconds