Effect of Healon5 and 4 other viscoelastic substances on intraocular pressure and endothelium after cataract surgery.

M P Holzer1, M R Tetz, G U Auffarth, R Welt, H E Völcker.   

Abstract

PURPOSE: To compare the ophthalmic viscoelastic device (OVD) Healon5 (sodium hyaluronate 2.3%) with 4 other commonly used OVDs during phacoemulsification and intraocular lens implantation in terms of influence on intraocular pressure (IOP) postoperatively and endothelial cells preoperatively and postoperatively.
SETTING: Department of Ophthalmology, Ruprecht-Karls-University Heidelberg, Germany.
METHODS: This clinical randomized prospective study, in which patients and observer were masked, comprised 81 eyes. Seventy-four eyes (mean patient age 71.2 years +/- 7.8 [SD]) completed all preoperative and 5 postoperative examinations. The OVDs used were OcuCoat and Celoftal (hydroxypropyl methylcellulose 2.0%), Viscoat (sodium hyaluronate 3.0%-chondroitin sulfate 4.0%), Healon GV (sodium hyaluronate 1.4%), and Healon5 (sodium hyaluronate 2.3%). Intraocular pressure was measured by standard Goldmann applanation tonometry preoperatively and 4 to 6 and 24 hours and 7, 30, and 90 days postoperatively. Endothelial cell counts were done preoperatively and 90 days postoperatively using a Pro/Koester WFSCM contact endothelial microscope. Exclusion criteria were IOP greater than 21 mm Hg at the preoperative examination, age younger than 40 years, significant corneal pathology, and a history or presence of uveitis or pseudoexfoliation syndrome.
RESULTS: All groups had increased IOP 4 hours postoperatively. The Healon5 group had the highest mean pressure (24.9 mm Hg) followed by the Viscoat group (23.6 mm Hg). The mean IOP in the other OVD groups was less than 22.1 mm Hg. These differences were not significant. Twenty-four hours postoperatively and at all subsequent examinations, mean IOP was below 20 mm Hg. The Healon5 group had the lowest mean endothelial cell loss (6.2%), significantly lower than in the other groups (P < .02).
CONCLUSION: With all 5 OVDs, endothelial cell loss was found, with the lowest in the Healon5 group, and IOP was increased 4 to 6 hours postoperatively. After 24 hours, no significant increases in IOP were noted.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11226784     DOI: 10.1016/s0886-3350(00)00568-x

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  18 in total

1.  Influence of anterior chamber depth, anterior chamber volume, axial length, and lens density on postoperative endothelial cell loss.

Authors:  Anna Reuschel; Holger Bogatsch; Nicole Oertel; Renate Wiedemann
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2015-03-01       Impact factor: 3.117

2.  Safety of hydroimplantation in cataract surgery in patients with pseudoexfoliation syndrome.

Authors:  Tevfik Oğurel; Reyhan Oğurel; Zafer Onaran; Kemal Örnek
Journal:  Int J Ophthalmol       Date:  2017-05-18       Impact factor: 1.779

3.  Complications and outcomes of descemet stripping automated endothelial keratoplasty with artisan aphakia intraocular lens implantation.

Authors:  Rong-Mei Peng; Yu-Xin Guo; Yuan Qiu; Yan-Sheng Hao; Jing Hong
Journal:  Int J Ophthalmol       Date:  2018-04-18       Impact factor: 1.779

4.  Application of sodium hyaluronate to the operation of suturing the corneal perforation.

Authors:  Tao Jiang; Jing Jiang; Yang Zhou; Gui-Qiu Zhao; Shan-Shan Yang; Hui Li; Li Shen
Journal:  Int J Ophthalmol       Date:  2010-09-18       Impact factor: 1.779

5.  Evaluation of the Rheological Properties, Preclinical Safety, and Clinical Effectiveness of a New Dispersive Ophthalmic Viscoelastic Device for Cataract Surgery.

Authors:  Claudia Palacio-Pastrana; Patricia Muñoz-Villegas; Fernando Dániel-Dorantes; Alejandra Sánchez-Ríos; Oscar Olvera-Montaño; Yareni I Martínez-Montoya; Juan D Quintana-Hau; Leopoldo M Baiza-Durán
Journal:  Med Devices (Auckl)       Date:  2022-08-24

6.  A clinical comparison between DisCoVisc and 2% hydroxypropylmethylcellulose in phacoemulsification: a fellow eye study.

Authors:  Rodrigo F Espíndola; Emerson F S Castro; Marcony R Santhiago; Newton Kara-Junior
Journal:  Clinics (Sao Paulo)       Date:  2012-09       Impact factor: 2.365

7.  Viscoat versus Visthesia during phacoemulsification cataract surgery: corneal and foveal changes.

Authors:  Marilita M Moschos; Irini P Chatziralli; Theodoros N Sergentanis
Journal:  BMC Ophthalmol       Date:  2011-04-29       Impact factor: 2.209

8.  Safety, efficacy, and intraoperative characteristics of DisCoVisc and Healon ophthalmic viscosurgical devices for cataract surgery.

Authors:  Satish S Modi; James A Davison; Tom Walters
Journal:  Clin Ophthalmol       Date:  2011-09-23

9.  Use of viscoelastic substance in ophthalmic surgery - focus on sodium hyaluronate.

Authors:  Tomomi Higashide; Kazuhisa Sugiyama
Journal:  Clin Ophthalmol       Date:  2008-03

10.  Comparative evaluation of efficacy and safety of ophthalmic viscosurgical devices in phacoemulsification [ISRCTN34957881].

Authors:  Rasik B Vajpayee; Kamna Verma; Rajesh Sinha; Jeewan S Titiyal; R M Pandey; Namrata Sharma
Journal:  BMC Ophthalmol       Date:  2005-07-15       Impact factor: 2.209

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.