Effect of peri-modiolar cochlear implant positioning on auditory nerve responses: a neural response telemetry study.

Stijn van Weert1, Robert J Stokroos, Mirçea M J G Rikers, Pim van Dijk.   

Abstract

CONCLUSIONS: There was no evidence that a reduction in current was needed for nerve stimulation as a result of modiolar placement of a cochlear implant. However, modiolus hugging did reduce the spread of excitation for the basal and apical electrodes. This improved stimulation selectivity may result in improved speech discrimination by implant users.
OBJECTIVE: To test the effect of modiolar placement of a cochlear implant on stimulation thresholds, and to confirm whether peri-modiolar electrode placement resulted in the hypothesized reduced spread of excitation.
MATERIAL AND METHODS: Auditory nerve responses were measured by means of neural response telemetry (NRT) in 14 subjects. All subjects received a Nucleus CI24R(CS) Contour implant. For each subject, the stimulation threshold and response growth rate were determined on all the odd-numbered electrodes, using a masker-and-probe paradigm. In addition, the spatial spread of excitation was measured on electrodes 1, 6, 11, 16 and 20, using a variable-masker paradigm. All NRT measurements were performed intra-operatively, both before and after peri-modular placement of the electrode by removal of surgical stylet.
RESULTS: Removal of the stylet had no significant effect on the threshold and growth rate of NRT responses. It caused a reduction in the spread of excitation for electrodes 1, 6 and 20, but not for electrodes 11 and 16.

Entities:  

Mesh:

Year:  2005        PMID: 16012034     DOI: 10.1080/00016480510028492

Source DB:  PubMed          Journal:  Acta Otolaryngol        ISSN: 0001-6489            Impact factor:   1.494


  8 in total

1.  Insertion trauma of a cochlear implant electrode array with Nitinol inlay.

Authors:  Thomas S Rau; Lenka Harbach; Nick Pawsey; Marcel Kluge; Peter Erfurt; Thomas Lenarz; Omid Majdani
Journal:  Eur Arch Otorhinolaryngol       Date:  2016-03-09       Impact factor: 2.503

2.  Relationship Between Electrode-to-Modiolus Distance and Current Levels for Adults With Cochlear Implants.

Authors:  Timothy J Davis; Dongqing Zhang; Rene H Gifford; Benoit M Dawant; Robert F Labadie; Jack H Noble
Journal:  Otol Neurotol       Date:  2016-01       Impact factor: 2.311

3.  Clinical evaluation of cochlear implant sound coding taking into account conjectural masking functions, MP3000™.

Authors:  Andreas Buechner; Andy Beynon; Witold Szyfter; Kazimierz Niemczyk; Ulrich Hoppe; Matthias Hey; Jan Brokx; Julie Eyles; Paul Van de Heyning; Gaetano Paludetti; Andrzej Zarowski; Nicola Quaranta; Thomas Wesarg; Joost Festen; Heidi Olze; Ingeborg Dhooge; Joachim Müller-Deile; Angel Ramos; Stephane Roman; Jean-Pierre Piron; Domenico Cuda; Sandro Burdo; Wilko Grolman; Samantha Roux Vaillard; Alicia Huarte; Bruno Frachet; Constantine Morera; Luis Garcia-Ibáñez; Daniel Abels; Martin Walger; Jochen Müller-Mazotta; Carlo Antonio Leone; Bernard Meyer; Norbert Dillier; Thomas Steffens; André Gentine; Manuela Mazzoli; Gerben Rypkema; Matthijs Killian; Guido Smoorenburg
Journal:  Cochlear Implants Int       Date:  2011-11

Review 4.  Challenging aspects of contemporary cochlear implant electrode array design.

Authors:  Pavel Mistrík; Claude Jolly; Daniel Sieber; Ingeborg Hochmair
Journal:  World J Otorhinolaryngol Head Neck Surg       Date:  2018-03-30

5.  Simpler and effective radiological evaluations for modiolar proximity of a slim modiolar cochlear implant electrode.

Authors:  Sang-Yeon Lee; Jin Hee Han; Marge Carandang; Yun Jung Bae; Byung Yoon Choi
Journal:  Sci Rep       Date:  2020-10-19       Impact factor: 4.379

6.  Advances to electrode pullback in cochlear implant surgery.

Authors:  Ingo Todt; Dietmar Basta; Rainer Seidl; Arne Ernst
Journal:  ScientificWorldJournal       Date:  2012-10-22

7.  Modeling of Auditory Neuron Response Thresholds with Cochlear Implants.

Authors:  Frederic Venail; Thibault Mura; Mohamed Akkari; Caroline Mathiolon; Sophie Menjot de Champfleur; Jean Pierre Piron; Marielle Sicard; Françoise Sterkers-Artieres; Michel Mondain; Alain Uziel
Journal:  Biomed Res Int       Date:  2015-07-05       Impact factor: 3.411

8.  Effects of in vivo repositioning of slim modiolar electrodes on electrical thresholds and speech perception.

Authors:  Sang-Yeon Lee; Young Seok Kim; Hyung Dong Jo; Yoonjoong Kim; Marge Carandang; Gene Huh; Byung Yoon Choi
Journal:  Sci Rep       Date:  2021-07-23       Impact factor: 4.379

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.