Literature DB >> 16011774

Defects in ProTaper S1 instruments after clinical use: longitudinal examination.

B Peng1, Y Shen, G S P Cheung, T J Xia.   

Abstract

AIM: To evaluate defects in ProTaper shaping instrument S1 after a defined schedule of clinical use.
METHODOLOGY: Among all ProTaper files discarded from an endodontic clinic at a stomatological school in China over a period of 17 months, 122 S1 instruments were collected. They were ultrasonically cleaned, autoclaved and then examined in the laboratory. Any instrument separation was noted; the average length involved was determined. The 0.5 mm region on either side of this length of discarded but intact instruments was examined circumferentially by scanning electron microscope. The region adjacent to the broken end of the fractured instruments was also examined in the same way.
RESULTS: One specimen was lost during processing. Of the remaining 121 instruments, 27 were separated with a mean fracture length of 3.67 mm from the tip. Of these, two files showed macroscopically torsional fracture and the others showed signs of flexural fatigue. Only one file that had not fractured showed visible unwinding defects. Examination of the 3.1-4.1 mm region of other unfractured instruments revealed the presence of microcracks, surface debris, pitting and/or wear of their cutting edges. Some debris particles seemed to have been trapped in crack-like structures.
CONCLUSIONS: Multiple use of ProTaper S1 pre-disposed the instrument to microcrack formation and wear of the cutting edges. There was a low prevalence of plastic deformation and most ProTaper S1 instruments failed without discernible sign of unwinding of the flutes. Further studies should address the mode of failure and the role of debris particles in the fracture mechanism.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16011774     DOI: 10.1111/j.1365-2591.2005.00991.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  17 in total

1.  Cutting efficiency of nickel-titanium rotary and reciprocating instruments after prolonged use.

Authors:  Gianluca Gambarini; Alessio Giansiracusa Rubini; Giampaolo Sannino; Gianni Di Giorgio; Fabrizio Di Giorgio; Lucila Piasecki; Dina Al-Sudani; Gianluca Plotino; Luca Testarelli
Journal:  Odontology       Date:  2014-11-30       Impact factor: 2.634

2.  Cyclic fatigue resistance of four nickel-titanium rotary instruments: a comparative study.

Authors:  Eugenio Pedullà; Gianluca Plotino; Nicola Maria Grande; Alfio Pappalardo; Ernesto Rapisarda
Journal:  Ann Stomatol (Roma)       Date:  2012-08-09

3.  Induction of osteogenic differentiation of adipose derived stem cells by microstructured nitinol actuator-mediated mechanical stress.

Authors:  Sarah Strauß; Sonja Dudziak; Ronny Hagemann; Stephan Barcikowski; Malte Fliess; Meir Israelowitz; Dietmar Kracht; Jörn W Kuhbier; Christine Radtke; Kerstin Reimers; Peter M Vogt
Journal:  PLoS One       Date:  2012-12-07       Impact factor: 3.240

4.  Comparative evaluation of surface changes in four Ni-Ti instruments with successive uses - An SEM study.

Authors:  N Subha; Vimal K Sikri
Journal:  J Conserv Dent       Date:  2011-07

5.  A comparative evaluation of cyclic fatigue resistance for different endodontic NiTi rotary files: An in-vitro study.

Authors:  Akanksha Bhatt; Balakrishnan Rajkumar
Journal:  J Oral Biol Craniofac Res       Date:  2019-01-02

6.  Failure of ProTaper rotary Ni-Ti instruments used by undergraduate students.

Authors:  Marc García-Font; Fernando Duran-Sindreu; Sergio Morello-Castro; Montserrat Mercade-Bellido; Rufino Bueno-Martínez; Miguel Roig-Cayón
Journal:  J Clin Exp Dent       Date:  2012-10-01

7.  The failures of root canal preparation with hand ProTaper.

Authors:  Marilena Bătăiosu; Oana Diaconu; Iren Moraru; C Dăguci; Mihaela Tuculină; Luminiţa Dăguci; Lelia Gheorghiţă
Journal:  Curr Health Sci J       Date:  2012-09-19

8.  Comparison of Cyclic Fatigue Resistance of Five Nickel Titanium Rotary File Systems with Different Manufacturing Techniques.

Authors:  Mohsen Aminsobhani; Naghmeh Meraji; Ehsan Sadri
Journal:  J Dent (Tehran)       Date:  2015-09

9.  The Effect of Autoclave Sterilization on Resistance to Cyclic Fatigue of Hero Endodontic File #642 (6%) at Two Artificial Curvature.

Authors:  Masoud Khabiri; Maziar Ebrahimi; Mohammad Reza Saei
Journal:  J Dent (Shiraz)       Date:  2017-12

10.  An analysis of cyclic fatigue resistance of reciprocating instruments in different canal curvatures after immersion in sodium hypochlorite and autoclaving: An in vitro study.

Authors:  C Champa; V Divya; A Srirekha; Rupali Karale; Ashwija Shetty; Pushpalatha Sadashiva
Journal:  J Conserv Dent       Date:  2017 May-Jun
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.