Literature DB >> 15993727

Attenuated pediatric electrode pads for automated external defibrillator use in children.

Dianne L Atkins1, Dawn B Jorgenson.   

Abstract

OBJECTIVE: This post-market, observational study is intended to evaluate reported uses of pediatric pads that reduce the energy delivered by some adult automated external defibrillators (AEDs) so that they may be used with pediatric patients.
METHODS: Users of the pediatric pads were asked to report any use of the pads, even if no shock was delivered and to provide detailed information about the event, caregiver and the patient.
RESULTS: Reports of the use of pediatric pads have been received and confirmed for 27 patients, age range 0 days to 23 years, median 2 years. Ventricular fibrillation (VF) was reported in eight cases, age range 4.5 months to 10 years, median 3 years. Shocks were delivered to all VF patients, the average shock number was 1.9, range 1-4. All patients had termination of VF, were admitted to the hospital and five survived to hospital discharge. Non-shockable rhythms were reported in 16 patients, and the AED appropriately did not advise a shock. Eleven of these patients had asystole or PEA as their initial rhythm and did not survive to hospital discharge. One report contained no additional information other than that the patient did not survive, and in two other reports, the pads were not applied to patients.
CONCLUSIONS: Voluntary reports of the use of attenuated pediatric defibrillation pads indicate the devices performed appropriately. All eight VF patients had termination of VF and five survived to hospital discharge. These data support the rapid deployment of AEDs for young children as well as adolescents and adults. Since the pediatric pads are available and deliver an appropriate dose for children, their use should be strongly encouraged.

Entities:  

Mesh:

Year:  2005        PMID: 15993727     DOI: 10.1016/j.resuscitation.2004.12.025

Source DB:  PubMed          Journal:  Resuscitation        ISSN: 0300-9572            Impact factor:   5.262


  12 in total

1.  Attenuating the defibrillation dosage decreases postresuscitation myocardial dysfunction in a swine model of pediatric ventricular fibrillation.

Authors:  Marc D Berg; Isabelle L Banville; Fred W Chapman; Robert G Walker; Mohammed A Gaballa; Ronald W Hilwig; Ricardo A Samson; Karl B Kern; Robert A Berg
Journal:  Pediatr Crit Care Med       Date:  2008-07       Impact factor: 3.624

2.  ECG screening of newborns to avoid sudden infant death syndrome?

Authors:  I M van Langen; A A M Wilde
Journal:  Neth Heart J       Date:  2006-02       Impact factor: 2.380

3.  Minimum equipment guidelines for paediatric prehospital care.

Authors:  A Cheng; D Hartfield
Journal:  Paediatr Child Health       Date:  2011-03       Impact factor: 2.253

4.  Targeted age, device deployment, and problems associated with pediatric defibrillation in pediatric prehospital emergency medical care settings in Japan.

Authors:  Noriyuki Kaku; Masahiko Nitta; Takashi Muguruma; Kohei Tsukahara; Emily Knaup; Nobuyuki Nosaka; Yuki Enomoto
Journal:  Acute Med Surg       Date:  2016-04-18

Review 5.  Part 13: pediatric basic life support: 2010 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care.

Authors:  Marc D Berg; Stephen M Schexnayder; Leon Chameides; Mark Terry; Aaron Donoghue; Robert W Hickey; Robert A Berg; Robert M Sutton; Mary Fran Hazinski
Journal:  Circulation       Date:  2010-11-02       Impact factor: 29.690

6.  Pediatric basic life support: 2010 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care.

Authors:  Marc D Berg; Stephen M Schexnayder; Leon Chameides; Mark Terry; Aaron Donoghue; Robert W Hickey; Robert A Berg; Robert M Sutton; Mary Fran Hazinski
Journal:  Pediatrics       Date:  2010-10-18       Impact factor: 7.124

Review 7.  A practical approach to paediatric emergencies in the radiology department.

Authors:  Nigel McBeth Turner
Journal:  Pediatr Radiol       Date:  2008-10-28

8.  Better outcome after pediatric resuscitation is still a dilemma.

Authors:  Sandeep Sahu; Kamal Kishore; Indu Lata
Journal:  J Emerg Trauma Shock       Date:  2010-07

Review 9.  Ventricular fibrillation and defibrillation.

Authors:  P Jones; N Lodé
Journal:  Arch Dis Child       Date:  2007-10       Impact factor: 3.791

10.  Defibrillation in children.

Authors:  Sarah E Haskell; Dianne L Atkins
Journal:  J Emerg Trauma Shock       Date:  2010-07
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.