Direct optical emission spectroscopy of liquid analytes using an electrolyte as a cathode discharge source (ELCAD) integrated on a micro-fluidic chip.

Gareth Jenkins1, Joachim Franzke, Andreas Manz.   

Abstract

Atomic emission detection of metallic species in aqueous solutions has been performed using a miniaturised plasma created within a planar, glass micro-fluidic chip. Detection was achieved using an Electrolyte as a Cathode Discharge source (ELCAD) in which the sample solution itself is used as the cathode for the discharge. To realise the ELCAD technique within a micro-fluidic device, a parallel liquid-gas flow was set up in a micro-channel and a glow discharge ignited between the flowing liquid sample surface and a metal wire anode. The detection of copper and sodium was achieved, using atmospheric pressure air as a carrier gas, by observation of atomic emission lines of copper at 324 nm, 327 nm, 511 nm, 515 nm and 522 nm and an atomic emission line of sodium at 589 nm using a commercially available miniaturised spectrometer. A total electrical power of less than 70 mW was required to sustain the discharge. A semi-quantitative, absolute detection limit of 17 nmol s(-1) was obtained for sodium with a sample flow rate of 100 microL min(-1) and an integration time of 100 ms in air at atmospheric pressure. The volume required for such detection is approximately 170 nL. Further analysis was performed with an Echelle spectrometer using both argon and air as a carrier gas. The geometry and flow rates used demonstrate the feasibility of integrating such micro-plasmas into other micro-fluidic devices, such as miniaturised CE devices, as a method of detection. The potential for using such micro-plasmas within highly portable miniaturised systems and mu-TAS devices is presented and discussed.

Entities:  

Year:  2005        PMID: 15970963     DOI: 10.1039/b501467k

Source DB:  PubMed          Journal:  Lab Chip        ISSN: 1473-0189            Impact factor:   6.799


  2 in total

1.  Microfluidic technology and its application in the point-of-care testing field.

Authors:  Yaping Xie; Lizhong Dai; Yijia Yang
Journal:  Biosens Bioelectron X       Date:  2022-01-20

2.  Electrically-triggered micro-explosion in a graphene/SiO2/Si structure.

Authors:  Siyang Liu; Myungji Kim; Hong Koo Kim
Journal:  Sci Rep       Date:  2018-05-09       Impact factor: 4.379

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.