Double-crowned valved stents for off-pump mitral valve replacement.

Liang Ma1, Piergiorgio Tozzi, Christoph H Huber, Steven Taub, Gabrielle Gerelle, Ludwig K von Segesser.   

Abstract

OBJECTIVE: An animal model has been designed to assess the feasibility of off-pump mitral valve replacement using valved stents.
METHODS: Glutaraldehyde-preserved homograft was sutured inside a prosthetic tube (Dacron). Then, two self-expandable nitinol Z-stents were sutured on the external surface of the prosthesis in such a way to create two self-expanding crowns for fixation. In adult pigs and under general anesthesia, the left atrium was exposed through a left thoracotomy and atrio-ventricular roadmapping was performed with intravascular ultrasound (IVUS) and fluoroscopy. The double-crowned valved stents were loaded into a delivery sheath. The sheath was then introduced into the left atrium and the valved stents was deployed in mitral position in such a way that the part in between the two stents was at the level of the mitral annulus. Intracardiac Unltrasound (ICUS) was used to assess the valve function. Hemodynamic parameters were gathered as well. Animal survived for no more than 3h after the valve deployment and gross anatomy examination of the left heart was carried out.
RESULTS: The mean height of the valved stents was 29.4+/-0.2 mm, with an internal diameter of 20.4+/-1.0mm, and an external diameter of 25.5+/-0.8 mm. The procedure was successfully carried out in eight animals. In vivo evaluation showed a native mitral annulus diameter of 24.9+/-0.6 mm, and a mean mitral valve area of 421.4+/-17.5 mm2. ICUS showed a mild mitral regurgitation in three out of eight animals. Mean pressure gradient across the valved stents was 2.6+/-3.1 mmHg. Mean pressure gradient across the left ventricular outflow tract (LVOT) was 6.6+/-5.2 mmHg. The mean survival time was 97.5+/-56.3 min (survival time range was 40-180 min). One animal died due to the occlusion of the LVOT because of valved stents displacement. Postmortem evaluation confirmed correct positioning of the valved stent in the mitral position in seven out of eight animals. No atrial or ventricular lesions due to the valved stents were found.
CONCLUSIONS: Off-pump implantation of a self-expandable valved stent in the mitral position is technically feasible. Further studies will assess if this procedure is also feasible in humans.

Entities:  

Mesh:

Year:  2005        PMID: 15950482     DOI: 10.1016/j.ejcts.2004.12.068

Source DB:  PubMed          Journal:  Eur J Cardiothorac Surg        ISSN: 1010-7940            Impact factor:   4.191


  11 in total

1.  True mitral annulus diameter is underestimated by two-dimensional echocardiography as evidenced by real-time three-dimensional echocardiography and magnetic resonance imaging.

Authors:  Ashraf M Anwar; Osama I I Soliman; Folkert J ten Cate; Attila Nemes; Jackie S McGhie; Boudewijn J Krenning; Robert-Jan van Geuns; Tjebbe W Galema; Marcel L Geleijnse
Journal:  Int J Cardiovasc Imaging       Date:  2006-12-13       Impact factor: 2.357

2.  Transapical access closure: the TA PLUG device.

Authors:  Henriette Brinks; Fabian Nietlispach; Volkhard Göber; Lars Englberger; Peter Wenaweser; Bernhard Meier; Thierry Carrel; Christoph Huber
Journal:  Interact Cardiovasc Thorac Surg       Date:  2013-07-09

3.  Melody valve-in-ring procedure for mitral valve replacement: feasibility in four annuloplasty types.

Authors:  Norihiro Kondo; Takashi Shuto; Jeremy R McGarvey; Kevin J Koomalsingh; Manabu Takebe; Robert C Gorman; Joseph H Gorman; Matthew J Gillespie
Journal:  Ann Thorac Surg       Date:  2012-03       Impact factor: 4.330

Review 4.  Devices for mitral valve repair.

Authors:  Paolo Denti; Francesco Maisano; Ottavio Alfieri
Journal:  J Cardiovasc Transl Res       Date:  2014-01-23       Impact factor: 4.132

5.  Percutaneous transvenous Melody valve-in-ring procedure for mitral valve replacement.

Authors:  Takashi Shuto; Norihiro Kondo; Yoav Dori; Kevin J Koomalsingh; Andrew C Glatz; Jonathan J Rome; Joseph H Gorman; Robert C Gorman; Matthew J Gillespie
Journal:  J Am Coll Cardiol       Date:  2011-12-06       Impact factor: 24.094

Review 6.  Will catheter interventions replace surgery for valve abnormalities?

Authors:  Michael L O'Byrne; Matthew J Gillespie
Journal:  Curr Opin Cardiol       Date:  2014-01       Impact factor: 2.161

7.  Transapical mitral implantation of the Tiara bioprosthesis: pre-clinical results.

Authors:  Shmuel Banai; Stefan Verheye; Anson Cheung; Marc Schwartz; Alexei Marko; Randy Lane; E Marc Jolicoeur; Patrick Garceau; Simon Biner; Jean-Francois Tanguay; Elazer R Edelman; Christopher J White
Journal:  JACC Cardiovasc Interv       Date:  2014-02       Impact factor: 11.195

8.  Sutureless mitral valve replacement: initial steps toward a percutaneous procedure.

Authors:  Matthew J Gillespie; Masahito Minakawa; Masato Morita; Mathieu Vergnat; Kevin J Koomalsingh; J Daniel Robb; Norihiro Kondo; Takashi Shuto; Manabu Takebe; Toru Shimaoka; Jeremy R McGarvey; Robert C Gorman; Joseph H Gorman
Journal:  Ann Thorac Surg       Date:  2013-08       Impact factor: 4.330

9.  Novel In Vitro Test Systems and Insights for Transcatheter Mitral Valve Design, Part I: Paravalvular Leakage.

Authors:  Eric L Pierce; Vahid Sadri; Beatrice Ncho; Keshav Kohli; Siddhi Shah; Ajit P Yoganathan
Journal:  Ann Biomed Eng       Date:  2018-10-17       Impact factor: 3.934

Review 10.  A Technical Review of Minimally Invasive Mitral Valve Replacements.

Authors:  Georgia L Preston-Maher; Ryo Torii; Gaetano Burriesci
Journal:  Cardiovasc Eng Technol       Date:  2014-11-25       Impact factor: 2.495

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.