Literature DB >> 15934012

Reproducibility and validity of the DynaPort KneeTest.

Lidwine B Mokkink1, Caroline B Terwee, Rienk M A van der Slikke, Rob C van Lummel, Rob J Benink, Lex M Bouter, Henrica C W de Vet.   

Abstract

OBJECTIVE: To determine the reproducibility and validity of the DynaPort KneeTest, a performance-based test that measures quality of movement of patients undergoing total knee replacement (TKR).
METHODS: A total of 92 patients with osteoarthritis (OA) of the knee performed the KneeTest twice on the same day; 94 healthy controls performed the KneeTest once. During the test, 29 activities were performed with accelerometers attached to the body. Relevant functional parameters were extracted from the accelerometers. A selection of parameters was used to calculate activity scores, based on the ability of parameters to discriminate between patients and controls (regression analyses). Based on internal consistency analyses (Cronbach's alpha), redundant activities were removed. Four cluster scores and a total KneeScore were calculated from the remaining activity scores. Reproducibility and validity of the cluster scores and the total KneeScore 2 were assessed.
RESULTS: Based on internal consistency analyses, the test was reduced to 23 activities. Inter- and intraobserver reliability using intraclass correlation coefficients were 0.90 (0.83-0.94) and 0.95 (0.83-0.98), respectively. Limits of inter- and intraobserver agreement were -8.3 to 11.3 and -4.2 to 9.0. Construct validity was confirmed by expected correlations with the Western Ontario and McMaster University Osteoarthritis Index physical functioning (0.55), Medical Outcomes Study Short Form-36 Health Survey physical functioning (0.62), and Knee Society Score function (0.64).
CONCLUSION: The KneeTest is a useful performance-based measure for research in patients with knee OA undergoing TKR, with good reliability and validity. Further research is required to improve its usefulness for clinical practice.

Entities:  

Mesh:

Year:  2005        PMID: 15934012     DOI: 10.1002/art.21167

Source DB:  PubMed          Journal:  Arthritis Rheum        ISSN: 0004-3591


  4 in total

1.  Functional improvement after unicompartmental knee replacement: a follow-up study with a performance based knee test.

Authors:  Lucas L A Kleijn; Wouter L W van Hemert; Will G H Meijers; Arnold D M Kester; Lukas Lisowski; Bernd Grimm; Ide C Heyligers
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2007-06-23       Impact factor: 4.342

2.  High correlation of the Oxford Knee Score with postoperative pain, but not with performance-based functioning.

Authors:  Ruud P van Hove; Richard M Brohet; Barend J van Royen; Peter A Nolte
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2015-03-29       Impact factor: 4.342

3.  Early functional outcome after subvastus or parapatellar approach in knee arthroplasty is comparable.

Authors:  Wouter L W van Hemert; Rachel Senden; Bernd Grimm; Matthijs J A van der Linde; Arno Lataster; Ide C Heyligers
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2010-10-17       Impact factor: 4.342

4.  Timed Up and Go test is predictive of Patient-Reported Outcomes Measurement Information System physical function in patients awaiting total knee arthroplasty.

Authors:  Deborah L Givens; Scott Eskildsen; Kaitlyn E Taylor; Richard A Faldowski; Daniel J Del Gaizo
Journal:  Arthroplast Today       Date:  2018-09-01
  4 in total