Literature DB >> 15933504

Facial nerve stimulation after cochlear implantation.

Jennifer L Smullen1, Marek Polak, Annelle V Hodges, Stacy B Payne, John E King, Fred F Telischi, Thomas J Balkany.   

Abstract

OBJECTIVES: This study was designed to compare the incidence and nature of facial nerve stimulation (FNS) in patients receiving cochlear implants (CI) manufactured by Cochlear Corporation, Advanced Bionics Corporation, and MedEl. STUDY
DESIGN: Retrospective chart review at a tertiary referral center.
METHODS: The charts of 600 patients who received CIs from 1993 to 2003 with at least 1 year of follow-up were reviewed for significant FNS (FNS on at least 1 channel at functional stimulation levels). Data collected included age, sex, etiology of deafness, device type, electrode, FNS onset after initial stimulation, number and location of electrode contacts causing FNS, and loudness level at which FSN occurred. Nucleus straight and perimodiolar electrodes were also compared.
RESULTS: Thirty-nine of 600 (6.5%) patients had FNS on at least one channel, (MedEl 3 of 43 [7.0%], Nucleus 29 of 440 [6.6%], and Clarion 7 of 117 [6.0%]). The incidence of FNS in Nucleus perimodiolar electrodes (16 of 250 [6.4%]) was similar to straight electrodes (13 of 190 (6.8%]), as was the mean number of electrodes causing FNS per patient (11 vs. 12). However, straight electrodes caused stimulation at significantly softer perceived loudness levels than perimodiolar electrodes (P < .0001).
CONCLUSIONS: In this large series of CI FNS, the overall incidence of FNS is consistent with previous reports. All devices had a similar incidence of FNS, but perimodiolar electrodes produced FNS only at significantly higher loudness levels than straight electrodes, making them preferable for patients at risk for FNS receiving Nucleus devices.

Entities:  

Mesh:

Year:  2005        PMID: 15933504     DOI: 10.1097/01.MLG.0000163100.37713.C6

Source DB:  PubMed          Journal:  Laryngoscope        ISSN: 0023-852X            Impact factor:   3.325


  12 in total

1.  Intra-Operative Neural Response Telemetry and Acoustic Reflex Assessment using an Advance-In-Stylet Technique and Modiolus-Hugging: A prospective cohort study.

Authors:  Abdulrahman Hagr
Journal:  Sultan Qaboos Univ Med J       Date:  2011-08-15

2.  Can Preoperative CT Scans Be Used to Predict Facial Nerve Stimulation Following CI?

Authors:  Jonathan L Hatch; Habib G Rizk; Michael W Moore; Elizabeth E Camposeo; Shaun A Nguyen; Paul R Lambert; Ted A Meyer; Theodore R McRackan
Journal:  Otol Neurotol       Date:  2017-09       Impact factor: 2.311

3.  Cochlear implantation with the nucleus slim modiolar electrode (CI532): a preliminary experience.

Authors:  Domenico Cuda; Alessandra Murri
Journal:  Eur Arch Otorhinolaryngol       Date:  2017-10-14       Impact factor: 2.503

Review 4.  Considerations for Fitting Cochlear Implants Bimodally and to the Single-Sided Deaf.

Authors:  Sabrina H Pieper; Noura Hamze; Stefan Brill; Sabine Hochmuth; Mats Exter; Marek Polak; Andreas Radeloff; Michael Buschermöhle; Mathias Dietz
Journal:  Trends Hear       Date:  2022 Jan-Dec       Impact factor: 3.496

5.  The pathologic basis of facial nerve stimulation in otosclerosis and multi-channel cochlear implantation.

Authors:  Mohammad Seyyedi; Barbara S Herrmann; Donald K Eddington; Joseph B Nadol
Journal:  Otol Neurotol       Date:  2013-12       Impact factor: 2.311

6.  Facial Nerve Stimulation as a Complication of Cochlear Implantation.

Authors:  Ihab Sefien; Soha Hamada
Journal:  Indian J Otolaryngol Head Neck Surg       Date:  2019-04-11

7.  Facial nerve stimulation after cochlear implantation: our experience.

Authors:  S Berrettini; De A Vito; L Bruschini; S Passetti; F Forli
Journal:  Acta Otorhinolaryngol Ital       Date:  2011-02       Impact factor: 2.124

8.  Comparison of Temporal Bone Parameters before Cochlear Implantation in Patients with and without Facial Nerve Stimulation.

Authors:  Venkatkrish Manohar Kasetty; Zachary Zimmerman; Sarah King; Mohammad Seyyedi
Journal:  J Audiol Otol       Date:  2019-09-24

9.  Detection of Extracochlear Electrodes in Cochlear Implants with Electric Field Imaging/Transimpedance Measurements: A Human Cadaver Study.

Authors:  Simone R de Rijk; Yu C Tam; Robert P Carlyon; Manohar L Bance
Journal:  Ear Hear       Date:  2020 Sep/Oct       Impact factor: 3.562

10.  Postoperative radiological assessment of the mastoid facial canal in cochlear implant patients in correlation with facial nerve stimulation.

Authors:  Iris Burck; Rania A Helal; Nagy N N Naguib; Nour-Eldin A Nour-Eldin; Jan-Erik Scholtz; Simon Martin; Martin Leinung; Silke Helbig; Timo Stöver; Annette Lehn; Thomas J Vogl
Journal:  Eur Radiol       Date:  2021-07-05       Impact factor: 5.315

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.