Literature DB >> 15922222

Evaluation of a continuous glucose monitoring system in cats with diabetes mellitus.

Jelena M E Ristic1, Michael E Herrtage, Sabine M M Walti-Lauger, Linda A Slater, David B Church, Lucy J Davison, Briän Catchpole.   

Abstract

A continuous glucose monitoring system (CGMS) was evaluated in 14 cats with naturally occurring diabetes mellitus. The device measures interstitial fluid glucose continuously, by means of a sensor placed in the subcutaneous tissue. All cats tolerated the device well and a trace was obtained on 15/16 occasions. There was good correlation between the CGMS values and blood glucose concentration measured using a glucometer (r=0.932, P<0.01). Limitations to the use of the CGMS are its working glucose range of 2.2-22.2 mmol/l (40-400 mg/dl) and the need for calibration with a blood glucose measurement at least every 12 h. When compared to a traditional blood glucose curve, the CGMS is minimally invasive, reduces the number of venepunctures necessary to assess the kinetics of insulin therapy in a patient and provides a truly continuous glucose curve.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15922222     DOI: 10.1016/j.jfms.2004.07.006

Source DB:  PubMed          Journal:  J Feline Med Surg        ISSN: 1098-612X            Impact factor:   2.015


  6 in total

1.  Recent advances in continuous glucose monitoring: biocompatibility of glucose sensors for implantation in subcutis.

Authors:  Peter H Kvist; Henrik E Jensen
Journal:  J Diabetes Sci Technol       Date:  2007-09

Review 2.  Monitoring methods for dogs and cats with diabetes mellitus.

Authors:  Audrey K Cook
Journal:  J Diabetes Sci Technol       Date:  2012-05-01

3.  Accuracy of a flash glucose monitoring system in dogs with diabetic ketoacidosis.

Authors:  Eleonora Malerba; Chiara Cattani; Francesca Del Baldo; Gaia Carotenuto; Sara Corradini; Stefania Golinelli; Ignazio Drudi; Federico Fracassi
Journal:  J Vet Intern Med       Date:  2019-11-14       Impact factor: 3.333

4.  Evaluation of the FreeStyle Libre, a flash glucose monitoring system, in client-owned cats with diabetes mellitus.

Authors:  Marieke Knies; Erik Teske; Hans Kooistra
Journal:  J Feline Med Surg       Date:  2022-06-28       Impact factor: 1.971

5.  Evaluation of glucose response to 3 types of insulin using a continuous glucose monitoring system in healthy alpacas.

Authors:  S R Byers; O M Beemer; A S Lear; R J Callan
Journal:  J Vet Intern Med       Date:  2014-06-25       Impact factor: 3.333

Review 6.  Pathophysiology and aetiology of hypoglycaemic crises.

Authors:  R K Morgan; Y Cortes; L Murphy
Journal:  J Small Anim Pract       Date:  2018-08-13       Impact factor: 1.522

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.