[Measurement of head penetration on digitalized radiographs: reproducibility and accuracy].

J Girard1, D Touraine, M Soenen, P Massin, P Laffargue, H Migaud.   

Abstract

PURPOSE OF THE STUDY: Wear of the acetabular component of total hip arthroplasty (THA) is incriminated as the cause of loosening and bone resorption. Consequently, an accurate evaluation of wear can contribute to the prediction of mechanical failure of the arthroplasty. Among the different methods proposed, digitalized imaging using a high-resolution scanner associated with data processing procedures appears to be a simple easily accessible technique. A system based on this concept has been introduced in orthopedics. To our knowledge, there has been no report on the system's reproducibility and accuracy.
MATERIAL AND METHODS: Thirty-nine radiography series for THA served as the basis for the evaluation of intra- and interobserver reproducibility. We evaluated the error induced by digitalization, the error induced by digitalized measurement, the accuracy of the measurements as a function of the material constituting the bearing, the intra- and interobserver reproducibility for repeated measures of THA wear (six observers and two observations). All measurements were done after digitalization and analyzed with a specimen designed software.
RESULTS: The inter and intra-observer coefficients of concordance were 0.6 and 0.58 respectively, i.e. moderate reproducibility. Depending on the prosthetic material, the error and accuracy of the system varied from 0.112 to 0.44 mm and 0.28 to 1.29 mm respectively. To obtain valid inter-observer reproducibility, the number of observers had to be limited to three (coefficient = 0.82). The type of implant had an influence on measurement error. The error was 0.342 for polyethylene cups and 0.118 for press-fit metal back cups. Likewise, for a metallic head measuring 22.2 mm, the error was 0.138 mm while for a ceramic head or metal head measuring 28 mm, the error was 0.28 mm and 0.112 mm respectively. DISCUSSION AND
CONCLUSION: The accuracy and error depend directly on the type of implant. The accuracy was better for metallic heads associated with metal-backed cups. To obtain satisfactory interclass correlation, the number of observers should be three. The proposed digitalized measurement system should be relatively accurate and reproducible. Its use can be recommended for the evaluation of wear after five years of follow-up, limiting the number of observers to three.

Entities:  

Mesh:

Year:  2005        PMID: 15908883     DOI: 10.1016/s0035-1040(05)84291-6

Source DB:  PubMed          Journal:  Rev Chir Orthop Reparatrice Appar Mot        ISSN: 0035-1040


  7 in total

1.  Does impact sport activity influence total hip arthroplasty durability?

Authors:  Matthieu Ollivier; Solenne Frey; Sebastien Parratte; Xavier Flecher; Jean-Noël Argenson
Journal:  Clin Orthop Relat Res       Date:  2012-11       Impact factor: 4.176

2.  No Benefit After THA Performed With Computer-assisted Cup Placement: 10-year Results of a Randomized Controlled Study.

Authors:  Sebastien Parratte; Matthieu Ollivier; Alexandre Lunebourg; Xavier Flecher; Jean-Noel A Argenson
Journal:  Clin Orthop Relat Res       Date:  2016-10       Impact factor: 4.176

3.  Can patients return to high-impact physical activities after hip resurfacing? A prospective study.

Authors:  Julien Girard; Bruno Miletic; Anthony Deny; Henri Migaud; Nicolas Fouilleron
Journal:  Int Orthop       Date:  2013-03-02       Impact factor: 3.075

4.  Changes of center of rotation and femoral offset in total hip arthroplasty.

Authors:  Jon A Bjarnason; Olav Reikeras
Journal:  Ann Transl Med       Date:  2015-12

5.  The John Charnley Award: The functional outcome of hip resurfacing and large-head THA is the same: a randomized, double-blind study.

Authors:  Martin Lavigne; Marc Therrien; Julie Nantel; Alain Roy; François Prince; Pascal-André Vendittoli
Journal:  Clin Orthop Relat Res       Date:  2009-06-20       Impact factor: 4.176

6.  The LP-ESP(®) lumbar disc prosthesis with 6 degrees of freedom: development and 7 years of clinical experience.

Authors:  Jean-Yves Lazennec; Alain Aaron; Adrien Brusson; Jean-Patrick Rakover; Marc-Antoine Rousseau
Journal:  Eur J Orthop Surg Traumatol       Date:  2013-01-11

7.  High 10-year survival rate with an anatomic cementless stem (SPS).

Authors:  Elhadi Sariali; Alexandre Mouttet; Philippe Mordasini; Yves Catonné
Journal:  Clin Orthop Relat Res       Date:  2012-03-09       Impact factor: 4.176

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.