Accuracy of automated CT angiography measurement of vascular diameter in phantoms: effect of size of display field of view, density of contrast medium, and wall thickness.

Shigeru Suzuki1, Shigeru Furui, Tatsuro Kaminaga.   

Abstract

OBJECTIVE: The objective of this study was to assess the effect of wall thickness, density of intravascular contrast material, and size of the display field of view on the accuracy of measurements of vascular diameter in phantoms yielded by automated software for CT angiography.
MATERIALS AND METHODS: Vascular models with three wall thicknesses (1.0, 0.8, and 0.5 mm) and an inner diameter of approximately 4 mm were filled with contrast material of three different densities (198, 270, and 350 H) and scanned with helical CT. Three sizes of display field of view (10, 15, and 20 cm) were used. We evaluated the measurement error of the automated software, which was defined as the difference between the diameter measurement of the automated software and the true inner diameter of the vascular model. Statistical analysis involved three-way analysis of variance with repeated measures.
RESULTS: There were significant differences in the measurement errors among the three wall thicknesses of the vascular model, three densities of intravascular contrast material, and three sizes of display field of view. The overall measurement errors progressively increased with larger sizes of display field of view (p < 0.01) and with lower densities of intravascular contrast material (p < 0.001). The measurement errors tended to progressively increase as the thickness of the wall of the vascular models increased.
CONCLUSION: The accuracy of the diameter measurements by automated software for CT angiography was affected by the size of the display field of view, intravascular density of the contrast material, and wall thickness of the vessel. It is necessary to consider the effects of these factors on the diameter measurements of small arteries.

Mesh:

Substances:

Year:  2005        PMID: 15908558     DOI: 10.2214/ajr.184.6.01841940

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  4 in total

1.  Effect of kVp on image quality and accuracy in coronary CT angiography according to patient body size: a phantom study.

Authors:  Sang Min Lee; Whal Lee; Jin Wook Chung; Eun-Ah Park; Jae Hyung Park
Journal:  Int J Cardiovasc Imaging       Date:  2013-11-02       Impact factor: 2.357

2.  Improved delineation of arteries in the posterior fossa of the brain by model-based iterative reconstruction in volume-rendered 3D CT angiography.

Authors:  H Machida; H Takeuchi; I Tanaka; R Fukui; Y Shen; E Ueno; S Suzuki; X-Z Lin
Journal:  AJNR Am J Neuroradiol       Date:  2013-01-04       Impact factor: 3.825

3.  Comparison of the second and third intercostal spaces regarding the use of internal mammary vessels as recipient vessels in DIEP flap breast reconstruction: An anatomical and clinical study.

Authors:  Ik Hyun Seong; Kyong-Je Woo
Journal:  Arch Plast Surg       Date:  2020-07-15

4.  Optimization of the scan protocol for the reduction of diaphragmatic motion artifacts depicted on CT angiography: a phantom study simulating pediatric patients with free breathing.

Authors:  Jee Hyun Baek; Whal Lee; Kee-Hyun Chang; Jin Wook Chung; Jae Hyung Park
Journal:  Korean J Radiol       Date:  2009-04-22       Impact factor: 3.500

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.