The effect of infection-control barriers on the light intensity of light-cure units and depth of cure of composite.

Nicholas A Hodson1, Stephen M Dunne, Caroline L Pankhurst.   

Abstract

AIM: Dental curing lights are vulnerable to contamination with oral fluids during routine intra-oral use. This controlled study aimed to evaluate whether or not disposable transparent barriers placed over the light-guide tip would affect light output intensity or the subsequent depth of cure of a composite restoration.
METHODS: The impact on light intensity emitted from high-, medium- and low-output light-cure units in the presence of two commercially available disposable infection-control barriers was evaluated against a no-barrier control. Power density measurements from the three intensity light-cure units were recorded with a radiometer, then converted to a digital image using an intra-oral camera and values determined using a commercial computer program. For each curing unit, the measurements were repeated on ten separate occasions with each barrier and the control. Depth of cure was evaluated using a scrape test in a natural tooth model.
RESULTS: At each level of light output, the two disposable barriers produced a significant reduction in the mean power density readings compared to the no-barrier control (P<0.005). The cure sleeve inhibited light output to a greater extent than either the cling film or the control (P<0.005). Only composite restorations light-activated by the high level unit demonstrated a small but significant decrease in the depth of cure compared to the control (P<0.05).
CONCLUSION: Placing disposable barriers over the light-guide tip reduced the light intensity from all three curing lights. There was no impact on depth of cure except for the high-output light, where a small decrease in cure depth was noted but this was not considered clinically significant. Disposable barriers can be recommended for use with light-cure lights.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15901434     DOI: 10.1308/1355761053695149

Source DB:  PubMed          Journal:  Prim Dent Care        ISSN: 1355-7610


  3 in total

1.  Evaluation of light-curing units in rural and urban areas.

Authors:  Maan M Alshaafi
Journal:  Saudi Dent J       Date:  2012-08-04

2.  Infection Control Barrier and Curing Time as Factors Affecting the Irradiance of Light-Cure Units.

Authors:  Reem A Ajaj; Hani M Nassar; Fatin A Hasanain
Journal:  J Int Soc Prev Community Dent       Date:  2018-11-29

3.  Effect of infection control barriers on the light output from a multi-peak light curing unit.

Authors:  Carlos José Soares; Stella Sueli Lourenço Braga; Maria Tereza Hordones Ribeiro; Richard Bengt Price
Journal:  J Dent       Date:  2020-10-19       Impact factor: 4.379

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.