Literature DB >> 15899656

Pitfall in the high-throughput quantification of whole blood cyclosporin A using liquid chromatography-tandem mass spectrometry.

Michael Vogeser1, Ute Spöhrer.   

Abstract

In a growing number of laboratories the technique of liquid chromatography-tandem mass spectrometry is used for the quantification of cyclosporin A in whole blood, employing cyclosporin D as the internal standard. Cyclosporin A is extensively metabolized in vivo; in liquid chromatography-tandem mass spectrometry respective metabolites can give rise to both parent and product ions that are isobaric with ions commonly used for the detection of cyclosporin A and cyclosporin D, respectively. In this article it is demonstrated that limited chromatography with co-elution of such metabolites together with cyclosporin A and cyclosporin D can lead to incorrect results.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15899656     DOI: 10.1515/CCLM.2005.072

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  3 in total

Review 1.  Usage and limitations of liquid chromatography-tandem mass spectrometry (LC-MS/MS) in clinical routine laboratories.

Authors:  Christoph Seger
Journal:  Wien Med Wochenschr       Date:  2012-11-06

2.  A rapid HPLC-MS/MS method for the simultaneous quantification of cyclosporine A, tacrolimus, sirolimus and everolimus in human blood samples.

Authors:  Christoph Seger; Karin Tentschert; Wolfgang Stöggl; Andrea Griesmacher; Steven L Ramsay
Journal:  Nat Protoc       Date:  2009       Impact factor: 13.491

3.  A new approach for the determination of immunosuppressive drugs using HPLC-MS/MS and Cs+ adducts.

Authors:  Patricia Kaiser; Theodorus Akerboom; William Graham Wood; Hans Reinauer
Journal:  Ger Med Sci       Date:  2006-01-18
  3 in total