B M Hoelper1, B Alessandri, A Heimann, R Behr, O Kempski. 1. Department of Neurosurgery, Klinikum Fulda, Academic Hospital of the Philips University Marburg, Fulda, Germany. hoelper@online.de
Abstract
BACKGROUND: Oxygen tension sensors have been used to monitor tissue oxygenation in human brain for several years. The working principals of the most frequently used sensors, the Licox (LX) and Neurotrend (NT), are different, and they have never been validated independently for correct measurement in vitro. Therefore, we tried to clarify if the two currently available sensors provide sufficient accuracy and stability. METHOD: 12 LX oxygen tension sensors and NT sensors were placed into a liquid-filled tonometer chamber. The solution was kept at 37 +/- 0.2 degrees C and equilibrated with five calibration gases containing different O(2)- and CO(2)-concentrations. After equilibration, readings were taken for each gas concentration (accuracy test). Afterwards, the sensors were left in 3% O(2) and 9% CO(2) and readings were taken after 24, 48, 72, 96 and 120 hours (drift test). Thereafter, a 90% response time test was performed transferring sensors from 1% to 5% oxygen concentration and back, using pre-equilibrated tonometers. FINDINGS: All Licox oxygen probes [12] were used for this study. Two of 14 Neurotrend sensors did not calibrate, revealing a failure rate of 14% for NT. Oxygen tension during the accuracy test was measured as follows: 1% O(2) (7.1 mmHg): LX 6.5 +/- 0.4, NT 5.3 +/- 2.3 mmHg, 2% O(2) (14.2 mmHg): LX 12.9 +/- 0.6, NT 12.1 +/- 2.2 mmHg, 3% O(2) (21.4 mmHg): LX 19.8 +/- 0.7, NT 19.4 +/- 2.4 mmHg, 5% O(2) (35.8 mmHg): LX 33.4 +/- 1.0 mmHg, NT 33.5 +/- 2.9 mmHg, 8% O(2) (57.0 mmHg): 53.8 +/- 1.5, NT 53.6 +/- 3.3 mmHg. After 120 hours in 3% O(2) (21 mmHg), LX measured 19.8 +/- 1.9 mmHg, NT 17.9 +/- 4.7 mmHg. 90% response time from 1% to 5%/5% to 1% oxygen concentration was 129 +/- 27/174 +/- 26 sec for LX, 55 +/- 19/98 +/- 39 sec for NT. CONCLUSIONS: Both systems are measuring oxygen tension sufficiently, but more accurately with LX probes. NT sensors read significantly lower pO(2) in 1% O(2) and show an increasing deviation with higher oxygen concentrations which was due to two of twelve probes. A slight drift towards lower oxygen tension readings for both sensors but more pronounced for the NT does not impair long-term use. NT measures pCO(2) and pH very accurately.
BACKGROUND:Oxygen tension sensors have been used to monitor tissue oxygenation in human brain for several years. The working principals of the most frequently used sensors, the Licox (LX) and Neurotrend (NT), are different, and they have never been validated independently for correct measurement in vitro. Therefore, we tried to clarify if the two currently available sensors provide sufficient accuracy and stability. METHOD: 12 LXoxygen tension sensors and NT sensors were placed into a liquid-filled tonometer chamber. The solution was kept at 37 +/- 0.2 degrees C and equilibrated with five calibration gases containing different O(2)- and CO(2)-concentrations. After equilibration, readings were taken for each gas concentration (accuracy test). Afterwards, the sensors were left in 3% O(2) and 9% CO(2) and readings were taken after 24, 48, 72, 96 and 120 hours (drift test). Thereafter, a 90% response time test was performed transferring sensors from 1% to 5% oxygen concentration and back, using pre-equilibrated tonometers. FINDINGS: All Licox oxygen probes [12] were used for this study. Two of 14 Neurotrend sensors did not calibrate, revealing a failure rate of 14% for NT. Oxygen tension during the accuracy test was measured as follows: 1% O(2) (7.1 mmHg): LX 6.5 +/- 0.4, NT 5.3 +/- 2.3 mmHg, 2% O(2) (14.2 mmHg): LX 12.9 +/- 0.6, NT 12.1 +/- 2.2 mmHg, 3% O(2) (21.4 mmHg): LX 19.8 +/- 0.7, NT 19.4 +/- 2.4 mmHg, 5% O(2) (35.8 mmHg): LX 33.4 +/- 1.0 mmHg, NT 33.5 +/- 2.9 mmHg, 8% O(2) (57.0 mmHg): 53.8 +/- 1.5, NT 53.6 +/- 3.3 mmHg. After 120 hours in 3% O(2) (21 mmHg), LX measured 19.8 +/- 1.9 mmHg, NT 17.9 +/- 4.7 mmHg. 90% response time from 1% to 5%/5% to 1% oxygen concentration was 129 +/- 27/174 +/- 26 sec for LX, 55 +/- 19/98 +/- 39 sec for NT. CONCLUSIONS: Both systems are measuring oxygen tension sufficiently, but more accurately with LX probes. NT sensors read significantly lower pO(2) in 1% O(2) and show an increasing deviation with higher oxygen concentrations which was due to two of twelve probes. A slight drift towards lower oxygen tension readings for both sensors but more pronounced for the NT does not impair long-term use. NT measures pCO(2) and pH very accurately.
Authors: Danila K Radolovich; Marek Czosnyka; Ivan Timofeev; Andrea Lavinio; Peter Hutchinson; Arun Gupta; John D Pickard; Peter Smielewski Journal: Neurocrit Care Date: 2009-01-30 Impact factor: 3.210
Authors: Marius M Mader; Anna Leidorf; Andreas Hecker; Axel Heimann; Petra S M Mayr; Oliver Kempski; Beat Alessandri; Gabriele Wöbker Journal: Neurocrit Care Date: 2018-10 Impact factor: 3.210
Authors: Stefan Boehme; Bastian Duenges; Klaus U Klein; Volker Hartwich; Beate Mayr; Jolanda Consiglio; James E Baumgardner; Klaus Markstaller; Reto Basciani; Andreas Vogt Journal: PLoS One Date: 2013-04-02 Impact factor: 3.240