Literature DB >> 15869683

Utility of echocardiographic tissue synchronization imaging to redirect left ventricular lead placement for improved cardiac resynchronization therapy.

Kaoru Dohi1, Matthew Suffoletto, Leonard Ganz, Marco Zenati, John Gorcsan.   

Abstract

An 80-year-old woman with severe symptomatic heart failure (ejection fraction of 13%), and left bundle branch block (QRS duration of 160 ms) underwent cardiac resynchronization therapy (CRT). She had significant baseline dyssynchrony with a septal to posterior wall delay of 160 ms by echocardiographic tissue synchronization imaging (TSI). Despite exhaustive efforts, a stable posterior-lateral coronary vein lead position could not be achieved with the standard percutaneous approach, resulting in anterior coronary vein lead placement. This resulted in no improvement in the patient's symptoms or ventricular function. Follow-up TSI revealed earlier activation of the anteroseptal site and worsened dyssynchrony with septal to posterior wall delay of now 290 ms. This information prompted surgical revision of the left ventricular (LV) lead position via limited thoracotomy and posterior-lateral epicardial lead implantation. Pacing at the new lead site resulted in a 30% increase in stroke volume and symptomatic improvement. TSI in this case redirected lead position in a clinical nonresponder, resulting in a favorable response to CRT.

Entities:  

Mesh:

Year:  2005        PMID: 15869683     DOI: 10.1111/j.1540-8159.2005.40056.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  5 in total

1.  Cardiac resynchronization therapy: the MGH experience.

Authors:  Jagmeet P Singh; Jeremy N Ruskin
Journal:  Ann Noninvasive Electrocardiol       Date:  2005-10       Impact factor: 1.468

Review 2.  "Dialing-in" cardiac resynchronization therapy: overcoming constraints of the coronary venous anatomy.

Authors:  Jagmeet P Singh; E Kevin Heist; Jeremy N Ruskin; J Warren Harthorne
Journal:  J Interv Card Electrophysiol       Date:  2007-01-25       Impact factor: 1.900

3.  Impact of left ventricular lead position on the efficacy of cardiac resynchronisation therapy: a two-dimensional strain echocardiography study.

Authors:  Michael Becker; Andreas Franke; Ole A Breithardt; Christina Ocklenburg; Theresa Kaminski; Rafael Kramann; Christian Knackstedt; Christoph Stellbrink; Peter Hanrath; Patrick Schauerte; Rainer Hoffmann
Journal:  Heart       Date:  2007-02-19       Impact factor: 5.994

4.  Application of CRT-D in a Marfan syndrome patient with chronic heart failure accompanied by ventricular tachycardia and ventricular fibrillation.

Authors:  Hai-ying Xu; Yi-zhou Xu; Feng Ling; Zhong Yu; Jun Yang; Xu Duan; Bei Wang; Jin-yu Huang
Journal:  J Zhejiang Univ Sci B       Date:  2013-08       Impact factor: 3.066

5.  A modified echocardiographic protocol with intrinsic plausibility control to determine intraventricular asynchrony based on TDI and TSI.

Authors:  Henryk Dreger; Adrian C Borges; Bruno Ismer; Sebastian Schattke; Berthold Stegemann; Gert Baumann; Christoph Melzer
Journal:  Cardiovasc Ultrasound       Date:  2009-09-25       Impact factor: 2.062

  5 in total