Literature DB >> 15862772

Low uncertainty reverse isotope dilution ICP-MS applied to certifying an isotopically enriched Cd candidate reference material: a case study.

Ivan Tresl1, Christophe R Quétel.   

Abstract

An analytical method is presented based on reverse isotope dilution single detector inductively coupled plasma magnetic sector mass spectrometry (ID-ICP-SMS) and applied to the specific case of the certification of a (111)Cd enriched candidate Cd spike calibration material (nominal mass fraction 10 mg kg(-1) in 5% HNO3 solution). Uncertainty propagation was used as a tool for both determining the analytical approach and validating it. The robustness of close to "exact matching" reverse IDMS to correction of measured isotope intensities for multiplicative (mass discrimination) and (semi)additive effects (dead time, instrumental background, and isobaric interference) is discussed. The very low experimental relative standard deviation of the mean (0.08%) of eight replicate determinations indicated that all significant sources of uncertainty had probably been taken into account for the estimation of the final combined uncertainty statement (U(c) = 0.17%, k = 1). IRMM-621 was used as comparator. Uncertainties on IUPAC isotopic abundances of 111Cd and 112Cd, for the natural Cd solution involved between the two enriched materials, formed nearly 60% of U(c). The repeatability of the isotope ratio measurements contributed less than 10%. Correction for procedural blank necessitated somewhat unusual calculations (potential contamination of an enriched material with natural Cd). The procedure also involved a quadrupole based ICP-MS judged to be appropriate for the characterization of the isotopic composition. For comparison purposes, direct IDMS results are simulated using identical experimental input data. Finally, a significant background signal in the 106-116 mass region, observed only with the magnetic sector instrument, was attributed to argon based isobaric interferences.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15862772     DOI: 10.1016/j.jasms.2005.01.022

Source DB:  PubMed          Journal:  J Am Soc Mass Spectrom        ISSN: 1044-0305            Impact factor:   3.109


  4 in total

1.  Comparative performance study of ICP mass spectrometers by means of U "isotopic measurements".

Authors:  C R Quétel; J Vogl; T Prohaska; S Nelms; P D Taylor; P De Bièvre
Journal:  Fresenius J Anal Chem       Date:  2000 Sep-Oct

2.  Isotope dilution inductively coupled plasma mass spectrometry (ID ICP-MS) for the certification of lead and cadmium in environmental standard reference materials.

Authors:  K E Murphy; E S Beary; M S Rearick; R D Vocke
Journal:  Fresenius J Anal Chem       Date:  2000-10

3.  SI-traceable certification of the amount content of cadium below the ng g(-1) level in blood samples by isotope dilution ICP-MS applied as a primary method of measurement.

Authors:  J Diemer; J Vogl; C R Quétel; T Linsinger; P D Taylor; A Lamberty; J Pauwels
Journal:  Fresenius J Anal Chem       Date:  2001-07

4.  Liquid-liquid extraction of cadmium by sodium diethyldithiocarbamate from biological matrixes for isotope dilution inductively coupled plasma mass spectrometry.

Authors:  R A Vanderpool; W T Buckley
Journal:  Anal Chem       Date:  1999-02-01       Impact factor: 6.986

  4 in total
  1 in total

1.  Two-dimensional heart-cut LC-LC improves accuracy of exact-matching double isotope dilution mass spectrometry measurements of aflatoxin B1 in cereal-based baby food, maize, and maize-based feed.

Authors:  Andreas Breidbach; Franz Ulberth
Journal:  Anal Bioanal Chem       Date:  2015-04       Impact factor: 4.142

  1 in total