| Literature DB >> 15846770 |
Tony Stöcker1, Frank Schneider, Martina Klein, Ute Habel, Thilo Kellermann, Karl Zilles, N Jon Shah.
Abstract
Standard procedures to achieve quality assessment (QA) of functional magnetic resonance imaging (fMRI) data are of great importance. A standardized and fully automated procedure for QA is presented that allows for classification of data quality and the detection of artifacts by inspecting temporal variations. The application of the procedure on phantom measurements was used to check scanner and stimulation hardware performance. In vivo imaging data were checked efficiently for artifacts within the standard fMRI post-processing procedure by realignment. Standardized and routinely carried out QA is essential for extensive data amounts as collected in fMRI, especially in multicenter studies. Furthermore, for the comparison of two different groups, it is important to ensure that data quality is approximately equal to avoid possible misinterpretations. This is shown by example, and criteria to quantify differences of data quality between two groups are defined.Mesh:
Year: 2005 PMID: 15846770 PMCID: PMC6871722 DOI: 10.1002/hbm.20096
Source DB: PubMed Journal: Hum Brain Mapp ISSN: 1065-9471 Impact factor: 5.038