Quadrupole mass spectrometer operating in the electron-capture negative ion mode as detector for comprehensive two-dimensional gas chromatography.

P Korytár1, J Parera, P E G Leonards, J de Boer, U A Th Brinkman.   

Abstract

A recently introduced rapid-scanning quadrupole mass spectrometer (qMS) with an electron-capture negative ion (ECNI) option, the Perkin-Elmer Clarus 500, was tested as a detector for comprehensive two-dimensional gas chromatography (GC x GC). The parameters influencing the data acquisition rate in the scan mode, such as scan time and inter-scan delay, and in the selected ion monitoring mode, such as dwell time and inter-channel delay, were evaluated. In the scan mode, good-quality mass spectra covering a range of 300 Da can be obtained at an acquisition rate of 23 Hz; in selected ion monitoring, an acquisition rate of 90Hz can be achieved when monitoring a single ion. Compared with electron ionisation, the use of electron-capture negative ionisation causes no extra peak broadening. As applications, mixtures of polychlorinated n-alkanes (PCAs), polybrominated diphenyl ethers (PBDEs) and polychlorinated dibenzo-p-dioxins (PCDDs) and dibenzofurans (PCDFs) were analysed. The separation of PCAs based on their number of chlorine substituents was confirmed for the first time by using GC x GC-ECNI qMS in the scan mode and a significantly improved limit of detection was achieved for BDEs (10-150 fg injected) and CDD/Fs (10-700 fg injected) in the selected ion monitoring mode.

Entities:  

Mesh:

Year:  2005        PMID: 15844531     DOI: 10.1016/j.chroma.2004.11.045

Source DB:  PubMed          Journal:  J Chromatogr A        ISSN: 0021-9673            Impact factor:   4.759


  2 in total

1.  Highly sensitive and selective analysis of urinary steroids by comprehensive two-dimensional gas chromatography combined with positive chemical ionization quadrupole mass spectrometry.

Authors:  Ying Zhang; Herbert J Tobias; J Thomas Brenna
Journal:  Analyst       Date:  2012-05-18       Impact factor: 4.616

2.  Pulsed large volume injection gas chromatography coupled with electron-capture negative ionization quadrupole mass spectrometry for simultaneous determination of typical halogenated persistent organic pollutants.

Authors:  Yuli Zhao; Limin Yang; Qiuquan Wang
Journal:  J Am Soc Mass Spectrom       Date:  2007-04-29       Impact factor: 3.109

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.