P Korytár1, J Parera, P E G Leonards, J de Boer, U A Th Brinkman.
Abstract
A recently introduced rapid-scanning quadrupole mass spectrometer (qMS) with an electron-capture negative ion (ECNI) option, the Perkin-Elmer Clarus 500, was tested as a detector for comprehensive two-dimensional gas chromatography (GC x GC). The parameters influencing the data acquisition rate in the scan mode, such as scan time and inter-scan delay, and in the selected ion monitoring mode, such as dwell time and inter-channel delay, were evaluated. In the scan mode, good-quality mass spectra covering a range of 300 Da can be obtained at an acquisition rate of 23 Hz; in selected ion monitoring, an acquisition rate of 90Hz can be achieved when monitoring a single ion. Compared with electron ionisation, the use of electron-capture negative ionisation causes no extra peak broadening. As applications, mixtures of polychlorinated n-alkanes (PCAs), polybrominated diphenyl ethers (PBDEs) and polychlorinated dibenzo-p-dioxins (PCDDs) and dibenzofurans (PCDFs) were analysed. The separation of PCAs based on their number of chlorine substituents was confirmed for the first time by using GC x GC-ECNI qMS in the scan mode and a significantly improved limit of detection was achieved for BDEs (10-150 fg injected) and CDD/Fs (10-700 fg injected) in the selected ion monitoring mode.Entities:
Mesh:
Year: 2005 PMID: 15844531 DOI: 10.1016/j.chroma.2004.11.045
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759