Comprehensive two dimensional gas chromatography (GC × GC) provides greater separation space than conventional GC. Because of fast peak elution, a time of flight mass spectrometer (TOFMS) is the usual structure-specific detector of choice. The quantitative capabilities of a novel GC × GC fast quadrupole MS were investigated with electron ionization (EI), and CH(4) or NH(3) positive chemical ionization (PCI) for analysis of endogenous urinary steroids targeted in anti-doping tests. Average precisions for steroid quantitative analysis from replicate urine extractions were 6% (RSD) for EI and 8% for PCI-NH(3). The average limits of detection (LODs) calculated by quantification ions for 12 target steroids spiked into steroid-free urine matrix (SFUM) were 2.6 ng mL(-1) for EI, 1.3 ng mL(-1) for PCI-CH(4), and 0.3 ng mL(-1) for PCI-NH(3), all in mass scanning mode. The measured limits of quantification (LOQs) with full mass scan GC × GC-qMS were comparable with the LOQ values measured by one-dimensional GC-MS in selected ion monitoring (SIM) mode. PCI-NH(3) yields fewer fragments and greater (pseudo)molecular ion abundances than EI or PCI-CH(4). These data show that a benchtop GC × GC-qMS system has the sensitivity, specificity, and resolution to analyze urinary steroids at normal urine concentrations, and that PCI-NH(3), not currently available on most GC × GC-TOFMS instruments, is of particular value for generation of structure-specific ions.Entities:
Mesh:
Substances:
Year: 2012 PMID: 22606686 PMCID: PMC4029849 DOI: 10.1039/c2an35087d
Source DB: PubMed Journal: Analyst ISSN: 0003-2654 Impact factor: 4.616