Literature DB >> 15843731

Three-dimensional dose verification of the clinical application of gamma knife stereotactic radiosurgery using polymer gel and MRI.

P Papagiannis1, P Karaiskos, M Kozicki, J M Rosiak, L Sakelliou, P Sandilos, I Seimenis, M Torrens.   

Abstract

This work seeks to verify multi-shot clinical applications of stereotactic radiosurgery with a Leksell Gamma Knife model C unit employing a polymer gel-MRI based experimental procedure, which has already been shown to be capable of verifying the precision and accuracy of dose delivery in single-shot gamma knife applications. The treatment plan studied in the present work resembles a clinical treatment case of pituitary adenoma using four 8 mm and one 14 mm collimator helmet shots to deliver a prescription dose of 15 Gy to the 50% isodose line (30 Gy maximum dose). For the experimental dose verification of the treatment plan, the same criteria as those used in the clinical treatment planning evaluation were employed. These included comparison of measured and GammaPlan calculated data, in terms of percentage isodose contours on axial, coronal and sagittal planes, as well as 3D plan evaluation criteria such as dose-volume histograms for the target volume, target coverage and conformity indices. Measured percentage isodose contours compared favourably with calculated ones despite individual point fluctuations at low dose contours (e.g., 20%) mainly due to the effect of T2 measurement uncertainty on dose resolution. Dose-volume histogram data were also found in a good agreement while the experimental results for the percentage target coverage and conformity index were 94% and 1.17 relative to corresponding GammaPlan calculations of 96% and 1.12, respectively. Overall, polymer gel results verified the planned dose distribution within experimental uncertainties and uncertainty related to the digitization process of selected GammaPlan output data.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15843731     DOI: 10.1088/0031-9155/50/9/004

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  9 in total

1.  Commissioning and benchmarking a 3D dosimetry system for clinical use.

Authors:  Andrew Thomas; Joseph Newton; John Adamovics; Mark Oldham
Journal:  Med Phys       Date:  2011-08       Impact factor: 4.071

Review 2.  Polymer gel dosimetry.

Authors:  C Baldock; Y De Deene; S Doran; G Ibbott; A Jirasek; M Lepage; K B McAuley; M Oldham; L J Schreiner
Journal:  Phys Med Biol       Date:  2010-02-11       Impact factor: 3.609

Review 3.  Image-guided convection-enhanced delivery platform in the treatment of neurological diseases.

Authors:  Massimo S Fiandaca; John R Forsayeth; Peter J Dickinson; Krystof S Bankiewicz
Journal:  Neurotherapeutics       Date:  2008-01       Impact factor: 7.620

4.  MAGAT gel and EBT2 film-based dosimetry for evaluating source plugging-based treatment plan in Gamma Knife stereotactic radiosurgery.

Authors:  Gopishankar Natanasabapathi; Vivekanandhan Subbiah; Shashank Sharad Kale; Goura Kishor Rath; S Senthilkumaran; Sanjay Thulkar; Vellaiyan Subramani; M A Laviraj; Raj Kishor Bisht; A K Mahapatra
Journal:  J Appl Clin Med Phys       Date:  2012-11-08       Impact factor: 2.102

5.  MRI-based polymer gel dosimetry for validating plans with multiple matrices in Gamma Knife stereotactic radiosurgery.

Authors:  N Gopishankar; Yoichi Watanabe; Vivekanandhan Subbiah
Journal:  J Appl Clin Med Phys       Date:  2011-01-31       Impact factor: 2.102

6.  Assessment of the accuracy and stability of frameless gamma knife radiosurgery.

Authors:  Hyun-Tai Chung; Woo-Yoon Park; Tae Hoon Kim; Yong Kyun Kim; Kook Jin Chun
Journal:  J Appl Clin Med Phys       Date:  2018-06-03       Impact factor: 2.102

Review 7.  Radiation Dosimetry by Use of Radiosensitive Hydrogels and Polymers: Mechanisms, State-of-the-Art and Perspective from 3D to 4D.

Authors:  Yves De Deene
Journal:  Gels       Date:  2022-09-19

8.  A novel method of estimating dose responses for polymer gels using texture analysis of scanning electron microscopy images.

Authors:  Cheng-Ting Shih; Jui-Ting Hsu; Rou-Ping Han; Bor-Tsung Hsieh; Shu-Jun Chang; Jay Wu
Journal:  PLoS One       Date:  2013-07-02       Impact factor: 3.240

9.  Mechanical and dose delivery accuracy evaluation in radiosurgery using polymer gels.

Authors:  Panagiotis Sandilos; Elias Tatsis; Lampros Vlachos; Pantelis Karaiskos; Evangelos Georgiou; Panagiotis Kipouros; Michael Torrens; Angelos Angelopoulos
Journal:  J Appl Clin Med Phys       Date:  2006-11-28       Impact factor: 2.102